Z-1765-2026Class II
Burlington Medical, Kilt Blockers
Burlington Medical LLC / initiated 2026-02-11 / Open, Classified
Burlington Medical, LLC began a recall of Burlington Medical, Kilt Blockers on . FDA classified it as Class II and gives the reason as "Potential for attenuation degradation over time, decreasing the lifespan". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1765-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Burlington Medical LLC
- Product
- Burlington Medical, Kilt Blockers.
- Product code
- IWO Apron, Protective
- Reason for recall
- Potential for attenuation degradation over time, decreasing the lifespan.
- FDA root cause
- Process change control
- Action
- Burlington Medical notified consignees on about 02/11/2026 via "URGENT - FIELD SAFETY NOTICE (FSN)" letter. Consignees were instructed to ensure all users of the specific Burlington Medical radiation protection garments made aware of this Recall Notice, review inventory for affected units, arrange for the return of those affected units for proper disposal, and complete and return the Recall Response Form. Consignees were also instructed to notify appropriate individuals within your organization and to any organization where the potentially affected devices were transferred or sold.
- Quantity in commerce
- 94 units
- Distribution
- Worldwide distribution. Nationwide distribution in the US and international distribution to Canada, Australia, Bangladesh, Chile, Hong Kong, India, Mexico, Malaysia, Panama, Singapore, Thailand, Taiwan, Uruguay, and Vietnam.
- Product codes and lots
- All serial numbers are affected that were manufactured with Xenolite 800 attenuation material from 01/16/2025 to 01/21/2026. 1. Model Number: R8KILTBLOCKER10X12; UDI-DI (Product Code): 00840331272727 (KILTBLOCKER8-R8KILTBLOCKER10X12). 2. Model Number: R8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011293 (KILTBLOCKER8-R8KILTBLOCKER12X12). 3. Model Number: R8KILTBLOCKER18X20; UDI-DI (Product Code): 00840331272741 (KILTBLOCKER8-R8KILTBLOCKER18X20). 4. Model Number: R8KILTBLOCKER20X24; UDI-DI (Product Code): 00840331272758 (KILTBLOCKER8-R8KILTBLOCKER20X24). 5. Model Number: R8KILTBLOCKER24X24; UDI-DI (Product Code): 00840331272765 (KILTBLOCKER8-R8KILTBLOCKER24X24). 6. Model Number: S8KILTBLOCKER10X10; UDI-DI (Product Code): 00840331272772 (KILTBLOCKER8-S8KILTBLOCKER10X10). 7. Model Number: S8KILTBLOCKER10X12; UDI-DI (Product Code): 00840331272789 (KILTBLOCKER8-S8KILTBLOCKER10X12). 8. Model Number: S8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011309 (KILTBLOCKER8-S8KILTBLOCKER12X12). 9. Model Number: S8KILTBLOCKER18X20; UDI-DI (Product Code): 00840331272802 (KILTBLOCKER8-S8KILTBLOCKER18X20). 10. Model Number: S8KILTBLOCKER20X24; UDI-DI (Product Code): 00840331272819 (KILTBLOCKER8-S8KILTBLOCKER20X24). 11. Model Number: S8KILTBLOCKER24X24; UDI-DI (Product Code): 00840331272826 (KILTBLOCKER8-S8KILTBLOCKER24X24). 12. Model Number: U8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011316 (KILTBLOCKER8-U8KILTBLOCKER12X12). 13. Model Number: U8KILTBLOCKER18X20; UDI-DI (Product Code): 00840331272864 (KILTBLOCKER8-U8KILTBLOCKER18X20). 14. Model Number: U8KILTBLOCKER20X24; UDI-DI (Product Code): 00840331272871 (KILTBLOCKER8-U8KILTBLOCKER20X24). 15. Model Number: U8KILTBLOCKER24X24; UDI-DI (Product Code): 00840331272888 (KILTBLOCKER8-U8KILTBLOCKER24X24).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1753-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1754-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1755-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1756-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1757-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1758-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1759-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1760-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1761-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1762-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1763-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1764-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1765-2026 | Class II | Ongoing | Voluntary: Firm initiated |