Z-1582-2026Class II
Convatec, EsteemBody Drainable Pouch REF:423643, 10-45mm
ConvaTec Limited / initiated 2026-02-16 / Open, Classified / root cause: manufacturing process
ConvaTec, Inc began a recall of Convatec, EsteemBody Drainable Pouch REF:423643, 10-45mm on . FDA classified it as Class II and gives the reason as "Drainable large pouch may leak due to manufacturing issue". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1582-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- ConvaTec Limited
- Product
- Convatec, EsteemBody Drainable Pouch REF:423643, 10-45mm
- Product code
- FON Bag, Drainage, With Adhesive, Ostomy
- Reason for recall
- Drainable large pouch may leak due to manufacturing issue.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On February XX, 2026 Convatec issued a updated Urgent Medical Device Recall Notification via E-Mail. The initial notification was issued February 16, 2026. Convatec asked consignees to take the following actions: 1. Review your personal supplies and determine if you are affected by this communication. 2. Discontinue use and dispose of the affected product. 3. Complete Response Form and return to Convatec within 30 days. 4. Acknowledgment of this action is critical. If you re not able to provide a written response liaise with your distributor or Convatec Customer Services can support. 5. Customer care team is available to assist you with any questions and provide product alternatives if needed. Please contact the regional Convatec Customer Services [email protected] Distributor: 1. Notify all downstream customers who may have received affected product. This letter may be used to support transmission of the communication. 2. For additional instructions reference customer notification
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Czech Republic, Denmark, Finland, Germany, Ireland, Italy, Japan, Netherlands, Norway, Poland, Slovakia, Sweden, Switzerland, UK.
- Product codes and lots
- UDI: 00768455221264 Lot: 5K00011,5K00714
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1582-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1583-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1584-2026 | Class II | Ongoing | Voluntary: Firm initiated |