Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1702-2026Class II

2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US

Synthes (USA) Products LLC / initiated 2026-02-25 / Open, Classified

Synthes (USA) Products LLC began a recall of 2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US on . FDA classified it as Class II and gives the reason as "A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1702-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes (USA) Products LLC
Product
2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
Product code
HWX Tap, Bone
Reason for recall
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
FDA root cause
Process change control
Action
Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.
Quantity in commerce
77 units
Distribution
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
Product codes and lots
Part Number: 03.424.124-US. GTIN: 10886982335944. Lot Number: 68900P9.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1702-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1702-2026Class IIOngoingVoluntary: Firm initiated
Z-1703-2026Class IIOngoingVoluntary: Firm initiated