Z-1703-2026Class II
2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US
Synthes (USA) Products LLC / initiated 2026-02-25 / Open, Classified
Synthes (USA) Products LLC began a recall of 2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US on . FDA classified it as Class II and gives the reason as "A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1703-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Synthes (USA) Products LLC
- Product
- 2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.
- Product code
- HWX Tap, Bone
- Reason for recall
- A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
- FDA root cause
- Process change control
- Action
- Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.
- Quantity in commerce
- 33 units
- Distribution
- US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
- Product codes and lots
- Part Number: 03.424.126-US. GTIN: 10886982335951. Lot Number: 68982P8.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1702-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||