LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the...
Magellan Diagnostics, Inc. began a recall of LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24... on . FDA classified it as Class II and gives the reason as "Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1895-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Magellan Diagnostics, Inc.
- Product
- LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24 each; 192 tests) " Treatment Reagent Tubes (8 containers of 24 each; 192 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
- Product code
- DOF Lead, Atomic Absorption
- Reason for recall
- Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On March 13, 2026, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. REQUIRED ACTIONS: This field action does NOT require removal of product from the field. The LeadCare Test Systems continue to perform as intended when used according to the product labeling and Instructions for Use. The regulatory status of all products noted above remains unchanged and the product supply uninterrupted. Technical and customer support will continue as normal. LEADCARE II: The only blood collection devices validated for use with the Magellan LeadCare II Blood Lead Test Kit are the capillary tubes provided in the LeadCare II Test Kit. No other collection device should be used for blood sample collection for testing with the LeadCare II System. LEADCARE PLUS AND ULTRA: End users must validate the use of micro-collection devices with LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems in accordance with their laboratory policies and procedures. REGARDING PREVIOUS RESULTS: Patients who have had elevated results when tested on the LeadCare Blood Lead Test Systems (LeadCare II, LeadCare Plus, and LeadCare Ultra) should have confirmatory testing performed in accordance with CDC or other applicable standards and guidelines.
- Quantity in commerce
- 284
- Distribution
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bhutan, Bolivia, Canada, Chile, Diego Garcia, Ecuador, Eritrea, Georgia, Ghana, Guam, Guatemala, Haiti, Hong Kong, India, Indonesia, Israel, Italy, Ivory Coast, Japan, Kyrgyzstan, Mexico, Nepal, Peru, Philippines, South Africa, South Korea, Spain, UAE, Uruguay, Vietnam.
- Product codes and lots
- Catalog Number: 70-8098; UDI-DI: 00850355006024; ALL lots and ALL sublots;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1893-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1894-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1895-2026 | Class II | Ongoing | Voluntary: Firm initiated |