Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2513-2026Class II

iLet Bionic Pancreas, REF: BB1001

Beta Bionics, Inc. / initiated 2025-09-23 / Open, Classified / root cause: software

Beta Bionics, Inc. began a recall of iLet Bionic Pancreas, REF: BB1001 on . FDA classified it as Class II and gives the reason as "The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2513-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Beta Bionics, Inc.
Product
iLet Bionic Pancreas, REF: BB1001
Product code
QFG Alternate Controller Enabled Insulin Infusion Pump
Reason for recall
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
Root cause tag
software
FDA root cause
Software design
Action
On 3/12/2026, recall notices were mailed and emailed to customers who were informed of the following: On September 23, 2025, Beta Bionics, Inc. issued an Urgent Medical Device Correction notification to customers informing them of the lock screen software issue and software upgrade. These problems are resolved in newer host software versions (3.3.7 and up). Please check your software version and update your software as soon as possible. Update your software 1. Download or update your iLet app on your phone. Need help! Read the iLet App User Guide. 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. 5. Complete the acknowledgement survey at https://sprw.io/stt-3tkjI or If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.
Quantity in commerce
4580
Distribution
US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
Product codes and lots
iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2513-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2513-2026Class IIOngoingVoluntary: Firm initiated