Z-1656-2026Class II

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Reflexion Medical, Inc. / initiated 2026-02-09 / Open, Classified / root cause: software

Reflexion Medical, Inc. began a recall of RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00 on . FDA classified it as Class II and gives the reason as "Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1656-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Reflexion Medical, Inc.
Product
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Product code
QVA Fludeoxyglucose F18-Guided Radiation Therapy System
Reason for recall
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Root cause tag
software
FDA root cause
Software change control
Action
On 2/9/2026, recall notices were emailed to customers who were informed of the following: How to recognize that the device may fail: Any patient treatment with a roll offset applied in a localization followed by a subsequent repeat localization within the same treatment session will be affected. For patient treatments that have already occurred, measuring the approximate distance off-axis (R) and the maximum roll offset value (., in radians) can provide an approximate displacement value (S) through a simple mathematical formula, which may be helpful in estimating the severity of the dose impact for the patient. Customers were asked to do the following: 1) Notify your clinical staff of the recall, including posting this notification on or near the system until this issue can be resolved. 2) Halt utilization of the Repeat Localization feature for all treatment and imaging sessions. a. Note: Standard Roll corrections may still be used on the system if they are only applied in the localization immediately prior to the treatment delivery or imaging session commencing, as those are unaffected by this issue. 3) If a Roll correction is applied in an initial Localization and a Repeat Localization is initiated, the user should end the treatment session without continuing to treatment delivery or PET imaging. The user should restart the treatment session as a new treatment session to properly apply the Roll correction in the initial Localization without performing a Repeat Localization afterwards. 4) Complete and return the acknowledgement form. Questions or additional assistance can be sent to the firm at [email protected]
Quantity in commerce
12
Distribution
US: CA, TX, PA, CT, NJ, OR, LA, OH
Product codes and lots
Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.