Z-2133-2026Class II

On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33

On-X Life Technologies Inc. / initiated 2026-02-01 / Open, Classified

On-X Life Technologies, Inc. began a recall of On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33 on . FDA classified it as Class II and gives the reason as "Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2133-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
On-X Life Technologies Inc.
Product
On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33
Product code
LWQ Heart-Valve, Mechanical
Reason for recall
Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications.
FDA root cause
Under Investigation by firm
Action
On 2/1/2026, the firm contacted their customer to quarantine affected devices.
Quantity in commerce
10
Distribution
International distribution to the country of South Korea.
Product codes and lots
UDI-DI: 00851788001303. Serial Numbers: A2013401, A2013402, A2013403, A2013404, A2013405, A2013407, A2013408, A2013409, A2013410, A2013411. Expiration: 2031Dec17

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2133-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2133-2026Class IIOngoingVoluntary: Firm initiated