Z-1852-2026Class II
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC
OLYMPUS Corporation of the Americas / initiated 2026-03-23 / Open, Classified / root cause: design
Olympus Corporation of the Americas began a recall of Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC on . FDA classified it as Class II and gives the reason as "Potential for detachment of a distal tip component of the device during use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1852-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Potential for detachment of a distal tip component of the device during use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were transferred or further distributed.
- Quantity in commerce
- 365 units
- Distribution
- International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
- Product codes and lots
- Model Number: TB2-0525FC. UDI-DI: 04953170440007. All Lot Numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1851-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1852-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1853-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1854-2026 | Class II | Ongoing | Voluntary: Firm initiated |