Z-1770-2026Class I

Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy)...

Boston Scientific Corporation / initiated 2026-03-19 / Open, Classified / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100... on . FDA classified it as Class I and gives the reason as "Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1770-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL Pacemaker, Model Number L101; 3. ESSENTIO SR SL MRI Pacemaker, Model Number L110; 4. ESSENTIO DR SL MRI Pacemaker, Model Number L111; 5. ESSENTIO DR EL Pacemaker, Model Number L121; 6. ESSENTIO DR EL MRI Pacemaker, Model Number L131; 7. PROPONENT SR SL Pacemaker, Model Number L200; 8. PROPONENT DR SL Pacemaker, Model Number L201; 9. PROPONENT DR (VDD) SL Pacemaker, Model Number L209; 10. PROPONENT SR SL Pacemaker, Model Number L210; 11. PROPONENT DR SL MRI Pacemaker, Model Number L211; 12. PROPONENT DR EL Pacemaker, Model Number L221; 13. PROPONENT DR EL MRI Pacemaker, Model Number L231*; 14. ACCOLADE SR SL Pacemaker, Model Number L300; 15. ACCOLADE DR SL Pacemaker, Model Number L301; 16. ACCOLADE SR SL MRI Pacemaker, Model Number L310; 17. ACCOLADE DR SL MRI Pacemaker, Model Number L311; 18. ACCOLADE DR EL Pacemaker, Model Number L321; 19. ACCOLADE DR EL MRI Pacemaker, Model Number L331*; 20. ALTRUA 2 SR SL Pacemaker, Model Number S701; 21. ALTRUA 2 DR SL Pacemaker, Model Number S702; 22. ALTRUA 2 DR EL Pacemaker, Model Number S722.
Product code
LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
Reason for recall
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Boston Scientific issued an Urgent Medical Device Correction notice to physicians, hospitals, and patients on 3/19/2026 via Federal Express or USPS certified mail. The notice explained the availability of the updated software, Brady Software maintenance release 6 (SMR6) for the ACCOLADE family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps), clinical and device longevity of the update, and recommendations for recommendations on the upgrade. In a separate notice, patients were advised to keep and attend all scheduled medical and device check appointments and talk to their doctor about the best steps for managing their pacemaker system. To confirm if the pacemaker or CRT-P is affected, please visit this website: www.BostonScientific.com/lookup. The language, as well as the model and serial numbers of the device, are required for the lookup which can be found on your patient ID card. Our Product Performance Resource Center at www.bostonscientific.com/ppr, includes information on this topic, a device lookup tool, and instructions for returning explanted products. If you have additional questions, or would like to report on a clinical event, please contact your Boston Scientific representative or Technical Services.
Quantity in commerce
718,456 units
Distribution
worldwide
Product codes and lots
The advisory population includes all models listed; however, the bounding differs by battery type: " All serialized DR-EL pacemakers and CRT-Ps from the ACCOLADE family are included in the advisory population. " ACCOLADE DR-SL and SR-SLs with a use-by-date (UBD) on or before 30 June 2025 are included in the advisory population. Model number alone will not precisely identify individual DR-SL or SR-SL devices in the advisory population. To determine if a device is affected, enter a model/serial into the device lookup tool at www.BostonScientific.com/lookup.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1770-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1770-2026Class IOngoingVoluntary: Firm initiated
Z-1771-2026Class IOngoingVoluntary: Firm initiated