Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13
Gentuity, LLC began a recall of Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13 on . FDA classified it as Class II and gives the reason as "The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1823-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Gentuity, LLC
- Product
- Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On March 23, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: The customers will be required to follow the updated instructions to correct for repeated frames per the revised Operator Manual to address longitudinal measurement accuracy until the updated version of the software that addresses this issue is installed on the console within 6 months.
- Quantity in commerce
- 184 units (28 US, 156 OUS)
- Distribution
- Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;
- Product codes and lots
- Model Number: G10-01; UDI-DI: 00859910007032; Lot/Serial Number: 23C0113, 22J0101, 20A0202, 23F0101, 22F0113, 22L0100, 26B0203, 26B0200, 26B0204, 26B0202, 26B0201, 25L0302, 22E0204, 22J0102, 24H0101, 24H0102, 23K0201, 20A0203, 22K0306, 20A0201, 25J0200, 23C0110, 22K0305, 25E0100, 22F0110, 22K0304, 22K0302, 24B0107; OUS Lot numbers: 21G1900, 21G1901, 21G1902, 21G1903, 21G1904, 21H1101, 21H1102, 21H1103, 21H1104, 21H1105, 22C2300, 22C2301, 22C3000, 22D2600, 22D2700, 22D2701, 22D2702, 22D2705, 22E0500, 22E0501, 22F2700, 22F2701, 22F2702, 22F2703, 22F2704, 22G2600, 22G2601, 22G2700, 22H2900, 22H2901, 24B2100, 24B2101, 24B2102, 24B2103, 24B2104, 24C1301, 24C1302, 24C1303, 24C1304, 24C1305, 24D1504, 24D1505, 24D1506, 24D1507, 24D1508, 24E0800, 24E0801, 24E0802, 24E0803, 24E0804, 24F2408, 24F2409, 24F2410, 24F2411, 24F2412, 24G2205, 24G2206, 24G2207, 24G2208, 24G2209, 24H1400, 24H1401, 24H1402, 24H1403, 24H1404, 24M1000, 22L3000, 22L3001, 23A1904, PDA23-001, 23C2202, 23C2203, 23E0900, 23E0901, 23E0902, 23F2704, 23F2705, 23G2601, 23G2602, 23G2603, 23G2604, 23G0700, 23G0701, 23H3000, 23H3001, 23H3002, 23H3003, 23M1102, 23M1103, 23M1104, 23M1301, 24B2105, 24B2106, 24C2200, 24C2201, 24H1900, 24H1901, 24H1902, 24J0500, 24J0501, 24J0502, 24M1101, 24M1102, 24M1103, 24M1104, 24M1105, 25C0320, 25C0321, 25C0322, 25C0323, 25C0324, 25F2500, 25F2501, 25F2502, 25G3103, 25G3104, 25G3105, 25J0800, 25J0801, 25J0802, 25J2300, 25K0909, 25K0910, 25L0306, 25L0307, 25M0203, 25M0201, 25M0202, 26B0310, 26B0311, 26B2300, 26B1900, 26B1901, 26B1902;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1823-2026 | Class II | Ongoing | Voluntary: Firm initiated |