Z-1931-2026Class II
LSL Healthcare, Infant Central Line Dressing Change Kit, Model/Catalog Number: 2925H
LSL Healthcare / initiated 2026-03-09 / Open, Classified / root cause: sterility
LSL Healthcare Inc. began a recall of LSL Healthcare, Infant Central Line Dressing Change Kit, Model/Catalog Number: 2925H on . FDA classified it as Class II and gives the reason as "Medical procedure kits were distributed containing Webcol Large Alcohol Prep Pads which were subsequently recalled for non-sterility". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1931-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- LSL Healthcare
- Product
- LSL Healthcare, Infant Central Line Dressing Change Kit, Model/Catalog Number: 2925H
- Product code
- OXQ Dressing Change Tray
- Reason for recall
- Medical procedure kits were distributed containing Webcol Large Alcohol Prep Pads which were subsequently recalled for non-sterility.
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- LSL Healthcare issued an UREGNT: Kit Component Recall to its consignees on 3/9/2026 via email. The notice explained the issue and requested the following: "1. Immediately stop using the affected kits in your facility. 2. Inspection all inventory locations within your facility and destroy all affected kits according to your facility's disposal process. If immediate destruction is not feasible, the products should be quarantined until proper disposal can occur. If disposal is not possible, please request a Return Merchandise Authorization (RMA) to return these kits as indicated.. 3. This recall should be executed at the distributor, hospital, and user levels. Identify all customers within your distribution network who may have purchased the affected products listed in this notification. 4. Complete the attached Response Form and return to the LSL contact provided on the form if you have any of the affected product in stock." For assistance with this process, please contact your LSL representative or LSL at LSL Healthcare, 6200 W. Howard Street, Nile, IL 60714, (888)225-5575, [email protected].
- Quantity in commerce
- 220 units
- Distribution
- US Nationwide distribution in the states of MD, IL, MN.
- Product codes and lots
- Lot Code: Model No. 2925H; UDI-DI 00661392052635; Lot#6B3022, Expiration Date 11/30/2026 GTIN: (01)00661392052635(10)6B3022(17)261130
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1928-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1929-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1930-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1931-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1932-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1933-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.