Diversatek Healthcare Viper 3-Stage Wire Guided Balloon Dilator, 18mm-19mm-20mm, REF 1205-20
Diversatek Healthcare began a recall of Diversatek Healthcare Viper 3-Stage Wire Guided Balloon Dilator, 18mm-19mm-20mm, REF 1205-20 on . FDA classified it as Class II and gives the reason as "The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1941-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Diversatek Healthcare Inc
- Product
- Diversatek Healthcare Viper 3-Stage Wire Guided Balloon Dilator, 18mm-19mm-20mm, REF 1205-20
- Product code
- KNQ Dilator, Esophageal
- Reason for recall
- The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Diversatek Healthcare issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 3/18/2026 via USPS. The notice explained the problem, risk to health, and requested the consignee discard any affected product. Complete and return the attached MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form and provide Signature Confirming Customer Receipt to Laura Boll, VP Regulatory Affairs & Quality at [email protected] or 414-755-4806 to receive a replacement order. For questions about this issue, please contact Laura Boll, Vice President - Regulatory Affairs & Quality, at 414-755-4806 or [email protected].
- Quantity in commerce
- 1383 units
- Distribution
- US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium
- Product codes and lots
- UDI/DI 00816734021804, Lot Numbers: 004947, 004950, 005311, 005397, 005484, 005574
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1938-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1939-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1940-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1941-2026 | Class II | Ongoing | Voluntary: Firm initiated |