Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1946-2026Class II

Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number...

Zimmer GmbH / initiated 2026-03-30 / Open, Classified / root cause: supplier

Zimmer, Inc. began a recall of Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product... on . FDA classified it as Class II and gives the reason as "Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1946-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Product code
JDZ Mixer, Cement, For Clinical Use
Reason for recall
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
On March 30, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Risk Managers were instructed to notify affected personnel and locate and quarantine affected product in inventory. A Zimmer Biomet sales representative may remove and return the affected product on your behalf, or the customer may choose to directly return all affected product from their facility. Surgeons were informed that there are no specific patient-monitoring instructions recommended beyond existing follow up schedules.
Quantity in commerce
66960
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.
Product codes and lots
Lot Code: Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301014(10)85390029 ; Individual UDI (01)00889024376564(17)301014(10)85390029 ; Lot Number 85390029 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301103(10)85682882 ; Individual UDI (01)00889024376564(17)301103(10)85682882 ; Lot Number 85682882 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)301001(10)85390028 ; Individual UDI (01)00889024376564(17)301001(10)85390028 ; Lot Number 85390028 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300422(10)82395239 ; Individual UDI (01)00889024376564(17)300422(10)82395239 ; Lot Number 82395239 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300423(10)82395240 ; Individual UDI (01)00889024376564(17)300423(10)82395240 ; Lot Number 82395240 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300929(10)85390027 ; Individual UDI (01)00889024376564(17)300929(10)85390027 ; Lot Number 85390027 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300924(10)85390026 ; Individual UDI (01)00889024376564(17)300924(10)85390026 ; Lot Number 85390026 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300415(10)82395238 ; Individual UDI (01)00889024376564(17)300415(10)82395238 ; Lot Number 82395238 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300922(10)85390025 ; Individual UDI (01)00889024376564(17)300922(10)85390025 ; Lot Number 85390025 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300826(10)85390022 ; Individual UDI (01)00889024376564(17)300826(10)85390022 ; Lot Number 85390022 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300918(10)85390024 ; Individual UDI (01)00889024376564(17)300918(10)85390024 ; Lot Number 85390024 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)300722(10)85390023 ; Individual UDI (01)00889024376564(17)300722(10)85390023 ; Lot Number 85390023 Model No. 00504901100 ; 20-Pack UDI (01)00889024379718(17)310122(10)86970970 ; Individual UDI (01)00889024376564(17)310122(10)86970970 ; Lot Number 86970970

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1946-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1946-2026Class IIOngoingVoluntary: Firm initiated