Z-2007-2026Class I

Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case

Medical Action Industries, Inc. 306 / initiated 2026-03-17 / Open, Classified

Medical Action Industries, Inc. 306 began a recall of Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case on . FDA classified it as Class I and gives the reason as "Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2007-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medical Action Industries, Inc. 306
Product
Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case
Product code
LRO General Surgery Tray
Reason for recall
Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"
FDA root cause
Unknown/Undetermined by firm
Action
On April 22, 2026, Medial Action Industries, Inc. (subsidiary of Owens & Minor) issued a "Urgent: Medical Device Recall Notification Update" via E-Mail to affected consignees. The updated notification provided clarity and and additional information for the recall strategy. Initial notification was issued on March 17, 2026. Medical Action Industries asked consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product. Convenience kits can still be used once properly labeled. 2. Add warning labels to affected Medical Action kits to ensure all users are aware of the actions below and identified in the Medline recall notice. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. 3. If your firm used the labeling template provided with the initial letter, please adhere the revised label over the existing template label. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by April 30, 2026. If you have any questions, contact Reg. Affairs Manager, Kitting, Global Products, by email: [email protected] or 470-280-4277.
Quantity in commerce
2030 kits
Distribution
US distribution to state of: AZ OUS: None
Product codes and lots
UDI:20809160490294/ Lot numbers: 342509, 346066, 338434, 339423, 337365, 335489, 336189, 336789

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2007-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2007-2026Class IOngoingVoluntary: Firm initiated