Z-2160-2026Class I

RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U...

Bolton Medical Inc. / initiated 2026-04-22 / Open, Classified / root cause: labeling and UDI

Bolton Medical Inc. began a recall of RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U... on . FDA classified it as Class I and gives the reason as "Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2160-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Bolton Medical Inc.
Product
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On June 15, 2026, Bolton Medical Inc. (subsidiary of Terumo Aortic) issued a 3rd iteration of the Urgent Medical Device Recall (Removal) notification via overnight mail. The updated communication contained a revised "Action to be taken" section as well as acknowledgment form. Initial notification was sent on April 22, 2026. and a 2nd iteration was issued on May 5, 2026, a 2nd iteration was issued updating the recall strategy to removal. Bolton Medical ask consignees to take the following actions: 1. Discontinue use of the Relay Pro immediately. 2. Review your Relay Pro Inventory and immediately isolate affected inventory to prevent future use. 3. If devices have been transferred to another facility, please provide them with a copy of this notification and instruct them to follow the actions in this section. 4. Complete the acknowledgment form indicating the number of units in your control to be returned to Terumo. The form is required even if you do not have product on hand. 6. Return affected product to the address below using the enclosed prepaid UPS return label and include a copy of Appendix 1 as the packing slip with your physical product return. For shipping assistance or additional labels, please call Sedgwick at 888-719-5830. Sedgwick, Inc.; Attn: Event 6732, 2670 Executive Dr., Suite A, Indianapolis, IN 46241.
Distribution
Worldwide - distribution in the US Territory of Puerto Rico and the countries of Argentina, Chile, Canada, Colombia, Mexico, United Kingdom.
Product codes and lots
All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)00843576150775; 28-N4-42-259-42U/ 42/ (01)00843576150904; 28-N4-44-114-44U/ 44/ (01)00843576150522; 28-N4-44-164-40U/ 44/ (01)00843576151017; 28-N4-44-164-44U/ 44/ (01)00843576150652; 28-N4-44-209-40U/ 44/ (01)00843576151116; 28-N4-44-209-44U/ 44/ (01)00843576150782; 28-N4-44-259-40U/ 44/ (01)00843576151215; 28-N4-44-259-44U/ 44/ (01)00843576150911; 28-N4-46-114-46U/ 46/ (01)00843576150539; 28-N4-46-164-42U/ 46/ (01)00843576151024; 28-N4-46-164-46U/ 46/ (01)00843576150669; 28-N4-46-209-42U/ 46/ (01)00843576151123; 28-N4-46-209-46U/ 46/ (01)00843576150799; 28-N4-46-259-42U/ 46/ (01)00843576151222

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2160-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2160-2026Class IOngoingVoluntary: Firm initiated