Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2149-2026Class II

Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300

GE Healthcare / initiated 2026-03-26 / Open, Classified / root cause: software

GE Medical Systems, LLC began a recall of Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300 on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2149-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Root cause tag
software
FDA root cause
Software design
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to [email protected] Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare
Quantity in commerce
25 units
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
UDI-DI: 00840682146173; Serial Numbers: AS10G2200136YC AS10G2200101YC AS10G2200068YC AS10G2200035YC AS10G2300093YC AS10G2400094YC AS10G2400079YC AS10G2100008YC AS10G2500002YC CBDWG2400019HM CBDWG2300081HM AS10G2400056YC AS10G2400081YC AS10G2400087YC CBDWG2400117HM CBDWG2400085HM CBDWG2500003HM CBDWG2400140HM CBDWG2400126HM AS10G2200085YC AS10G2300057YC CBDWG2500011HM CBDWG2300056HM CBDWG2300041HM AS10G2300041YC

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2149-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2149-2026Class IIOngoingVoluntary: Firm initiated
Z-2150-2026Class IIOngoingVoluntary: Firm initiated
Z-2151-2026Class IIOngoingVoluntary: Firm initiated
Z-2152-2026Class IIOngoingVoluntary: Firm initiated
Z-2153-2026Class IIOngoingVoluntary: Firm initiated
Z-2154-2026Class IIOngoingVoluntary: Firm initiated
Z-2155-2026Class IIOngoingVoluntary: Firm initiated