Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI...
ARROW INTERNATIONAL, LLC began a recall of Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI... on . FDA classified it as Class I and gives the reason as "16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2157-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-15552-VFIE; Arrow-Clark VectorFlow Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: REF: CS-15192-VFIM, CS-15232-VFIM, CS-15272-VFIM; Arrow-Clark VectorFlow Retrograde Chronic Hemodialysis Catheter, REF: ACS-15192-VF, ACS-15232-VF, ACS-15272-VF, ACS-15552-VF, CS-15192-VF, CS-15192-VFE, CS-15232-VF, CS-15232-VFE, CS-15272-VF, CS-15272-VFE, CS-15312-VF, CS-15312-VFE, CS-15422-VF, CS-15422-VFE, CS-15552-VF, CS-15552-VFE; Arrow-Clark VectorFlow Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: CS-15192-VFM, CS-15232-VFM, CS-15272-VFM
- Product code
- MSD Catheter, Hemodialysis, Implanted
- Reason for recall
- 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 4/10/2026, recall notices were mailed to customers who were informed of the following Arrow International LLC is recalling product due to the Merit Medical Systems, Inc. recall of the Merit 16F Dual-Valved Splitable Sheath Introducer component, which is included in impacted Arrow Hemodialysis Kits & Sets. Customers were asked to do the following: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Complete and return the Acknowledgement Form, see link: https://fca.myteleflex.com/en/recall/000605 3) This notice should be passed on to all persons who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. In addition, distributors were asked to do the following: - Provide a copy of this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. Should you require further information or support concerning this issue, please contact Customer Service via email ([email protected]), phone (1-866-396-2111), or FAX (1-855-419-8507).
- Quantity in commerce
- 92,827
- Distribution
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
- Product codes and lots
- REF/UDI-DI/Lots: ACS-15192-VFI/ 10801902126709/33F23G0040, 33F25C0689; ACS-15232-VFI/10801902126723/33F23G0045, 33F25C0678; ACS-15272-VFI/ 10801902126747/33F24G0638; CS-15192-VFI/10801902127188, 10801902193572/33F23E0579, 33F23G0042, 33F23K0170, 33F23L0933, 33F24E0271, 33F24G0166, 33F24J0086, 33F24L0064, 33F24M0224, 33F25C0652, 33F25D0052, 33F25D0700, 33F25E1081, 33F25F0015, 33F25F0388, 33F25G0702, 33F25H0785; CS-15192-VFIE/10801902199659/33F23D0853, 33F23E0773, 33F23G0044, 33F23G0532, 33F23H0611, 33F23K0169,33F23K1105,33F23L0934, 33F24A0912, 33F24A0913, 33F24C0058, 33F24D1001, 33F24G0167,33F24G0187, 33F24J0030, 33F24J0032, 33F24J0339, 33F24J0381, 33F24K0659, 33F24L0058, 33F24L0511, 33F24M0155, 33F24M0215, 33F24M0265, 33F25A0143, 33F25A0407, 33F25B0259, 33F25C0037, 33F25D0285, 33F25D0738, 33F25F0060, 33F25G0635; CS-15232-VFI/10801902194463/33F23C0864, 33F23E0578, 33F23E0832, 33F23G0535, 33F23G0840, 33F23H0612, 33F23K0144, 33F24A0914, 33F24C0079, 33F24E0269, 33F24J0042, 33F25B0256, 33F25C0014, 33F25C0709, 33F25C0650, 33F25E1085, 33F25E1130, 33F25F0009, 33F25G0023, 33F25G0450, 33F25H0341, 33F25J1014; CS-15232-VFIE/10801902195149/33F23D0855, 33F23E0774, 33F23G0047, 33F23G0048, 33F23G0049, 33F23K0145, 33F23K0175, 33F23L0938, 33F24A0915, 33F24C0069, 33F24G0169, 33F24J0026, 33F24J0027, 33F24J0028, 33F24J0084, 33F24K0700, 33F24L0178, 33F24M0005, 33F24M0214, 33F24M0219, 33F25C0646, 33F25C0665, 33F25D0097, 33F25E0358, 33F25E0434, 33F25F0012, 33F25F0275, 33F25F0328, 33F25G0069, 33F25H0293, 33F25H0822, 33F25K0235; CS-15272-VFI/10801902193640, 10801902127362/33F23E0871, 33F23F0742, 33F23H0981, 33F23K0181, 33F24A0306, 33F24C0084, 33F24J0062, 33F25D0098, 33F25E0307, 33F25E0332, 33F25E0354, 33F25F0058, 33F25F0245, 33F25J0659; CS-15272-VFIE/10801902195255/33F23E0886, 33F23G0537, 33F23H0982, 33F23K0180, 33F23L0946, 33F24A0922, 33F24C0065, 33F24D1011, 33F24E0686, 33F24E0805, 33F24K0508, 33F24M0223, 33F25A0410, 33F25C0009, 33F25G0717; CS-15312-VFI/10801902197778/33F23G0543, 33F23L0953, 33F24D1007, 33F24E0810, 33F24J0048, 33F25C0039, 33F25D0218, 33F25E0470, 33F25J0021; CS-15312-VFIE/10801902195491/33F23K0183, 33F23L0954, 33F24A0923, 33F24D1013, 33F25A0403, 33F25F0534, 33F25G0722; CS-15422-VFI/10801902193664/33F23G0549, 33F24E0815, 33F25E1074; CS-15422-VFIE/10801902197440/33F23G0550, 33F23K0178, 33F23L0958, 33F24A0919, 33F24C0073, 33F24E0816, 33F25F0303; CS-15552-VFI/10801902193671/33F23K0184, 33F24J0076, 33F25E1075; CS-15552-VFIE/10801902195163/33F23E0865, 33F23F0743, 33F24D0972, 33F24E0819, 33F24G0098, 33F24K0656, 33F25A0032; CS-15192-VFIM/10801902193589/33F24E0792, 33F25C0676, 33F25L0035; CS-15232-VFIM/10801902193619/33F23C0879, 33F23L0939, 33F24A0304, 33F24E0798, 33F24J0047; CS-15272-VFIM/10801902197761/33F23H1027; ACS-15192-VF/10801902126693/33F24G0637; ACS-15232-VF/10801902126716/33F23H1016; ACS-15272-VF/10801902126730/33F23H1024; ACS-15552-VF/10801902126792/33F23H1021; CS-15192-VF/20801902193593/33F23K0186, 33F23L0931, 33F24C0082, 33F24E0788, 33F24J0054, 33F24M0162, 33F24M0230, 33F25D0271; CS-15192-VFE/10801902195460/33F23E0776, 33F23F0728, 33F23G0531, 33F24A0917, 33F24C0063, 33F24D1003, 33F24E0789, 33F24J0039, 33F24L0069, 33F24M0156, 33F24M0159, 33F25C0683, 33F25D0107, 33F25D0220, 33F25E0343, 33F25F0273, 33F25F0462, 33F25G0444, 33F25H0820, 33F25K0384; CS-15232-VF/10801902193626/33F23C0873, 33F23J0546, 33F24A0294, 33F24C0071, 33R24G0168, 33F24J0044, 33F25D0089, 33F25E0333, 33F25E0427, 33F25E1095, 33F25F0404, 33F25H0294, 33F25J1009; CS-15232-VFE/10801902195132/33F23E0775, 33F23F0720, 33F23K0176, 33F23K0194, 33F23L0937, 33F24A0927, 33F24C0059, 33F24D1010, 33F24E0270, 33F24E0796, 33F24G0045, 33F24J0035, 33F24J0083, 33F24L0171, 33F24M0216, 33F25A0004, 33F25A0083, 33F25C0023, 33F25C0659, 33F25D0106, 33F25D0219, 33F25D0726, 33F25E0478, 33F25E1105, 33F25F0298, 33F25J1056, 33F25K0335, 33F25L0011; CS-15272-VF/10801902193657/33F23C0880, 33F23F0740, 33F23L0944, 33F24C0081, 33F24E0804, 33F24J0051, 33F24J0686, 33F24M0228, 33F25C0668, 33F25E0337, 33F25F0403, 33F25J0590; CS-15272-VFE/10801902195248/33F23E0777, 33F23F0732, 33F23H0980, 33F23K0172, 33F23L0945, 33F24A0918, 33F24C0064, 33F24H0044, 33F24H0255, 33F24M0237, 33F25C0049, 33F25E0349, 33F25E1132, 33F25G0637; CS-15312-VF/10801902127447/33F24G0060, 33F25D0216; CS-15312-VFE/10801902195484/33F23F0734, 33F23G0542, 33F23L0952, 33F24A0925, 33F24C0076, 33F24J0053, 33F24M0225, 33F25A0033, 33F25C0667, 33F25E0320, 33F25E0469, 33F25G0603; CS-15422-VF/10801902128666/33F23H1020, 33F25G0734; CS-15422-VFE/10801902197341/33F23G0548, 33F23H0986, 33F23L0957, 33F24A0929, 33F24J0483, 33F24M0229, 33F25C0681, 33F25E0440, 33F25F0250, 33F25G0484; CS-15552-VF/10801902128703/33F23H1012, 33F25C0674, 33F25J0588; CS-15552-VFE/10801902195156/33F23D0868, 33F23G0261, 33F23G0551, 33F23K0173, 33F24A0928, 33F24D1014, 33F24E0818, 33F24J0059, 33F24K0511, 33F25C0670, 33F25E0471, 33F25F0455, 33F25G0447; CS-15192-VFM/10801902193602/33F23K0189, 33F24C0078, 33F24J0049, 33F24M0243, 33F25C0672, 33F25E1077, 33F25F0418; CS-15232-VFM/10801902193633/33F23J0547, 33F24A0289, 33F24C0077, 33F24E0799, 33F24J0045, 33F24J0093, 33F25A0026, 33F25C0050, 33F25E0348, 33F25K0348; CS-15272-VFM/10801902127393/33F25C0687
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2156-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2157-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2158-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2159-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.