Z-2212-2026Class I

Pediatric care bed; Product Designation: KayserBett IDA

Kayserbetten GmbH & Co. KG / initiated 2026-05-04 / Open, Classified / root cause: design

KAYSERBETTEN GMBH & CO. KG began a recall of Pediatric care bed; Product Designation: KayserBett IDA on . FDA classified it as Class I and gives the reason as "If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2212-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Kayserbetten GmbH & Co. KG
Product
Pediatric care bed; Product Designation: KayserBett IDA;
Product code
FNL Bed, Ac-Powered Adjustable Hospital
Reason for recall
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
Root cause tag
design
FDA root cause
Device Design
Action
On May 4, 2026 Urgent Field Safety Notice letters were sent to customers. Actions to be taken: Please inform all users and operators of a KayserBett IDA that by implementing the following safety measures, the above-mentioned risk to patients or third parties can be eliminated during further use of the product. If the patient is unattended, " the sleeping platform must be lowered to its lowest position. " the adjustment functions on the hand control must be locked using a key. " the doors must be closed. KayserBetten is implementing a corrective action to replace the hand control on all affected KayserBett IDA beds in the U.S. field with a modified hand control featuring an automatic locking function (Auto-Lock). With the new hand control, the adjustment functions lock automatically after each use, eliminating the dependence on the caregiver manually locking the hand control with a key. The hand control replacement will be performed by an authorized service technician at no cost to the customer. Your U.S. distributor, Mobility Unlimited Inc. (DBA KayserBetten US), will contact you to schedule the replacement. Please ensure in your organization that all users of the above-mentioned product and other persons to be informed are aware of this Field Safety Notice. If you have supplied the products to third parties, please forward a copy of this information or inform the contact person listed below. Please keep this information at least until the hand control replacement has been completed.
Quantity in commerce
372 units
Distribution
Distribution US Nationwide and Canada.
Product codes and lots
Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819, 10820, 10821, 10822, 10823, 10824, 10825, 11186, 11367, 11368, 11369, 11370, 11371, 11372, 11797, 11798, 11799, 11800, 11801, 11802, 11803, 11804, 11805, 11806, 11807, 11808, 12247, 12248, 12249, 12250, 12251, 12252, 12253, 12254, 12255, 12256, 12257, 12397, 12398, 12399, 12782, 12783, 12963, 12964, 12965, 12966, 12967, 12968, 12969, 12970, 12971, 12972, 13486, 13487, 13488, 13489, 13490, 13491, 13492, 13493, 13494, 13495, 13496, 13497, 13912, 13913, 13914, 13915, 13916, 13917, 13918, 14942, 14944, 14952, 14953, 15912, 15913, 15914, 16403, 16404, 16405, 16406, 16407, 16408, 16409, 16410, 16411, 16984, 16985, 16986, 16987, 16988, 16989, 16990, 16991, 16992, 16993, 16994, 17577, 17578, 17579, 17580, 17581, 17582, 17583, 17584, 17585, 17586, 17587, 18494, 19112, 19113, 19114, 20151, 20152, 20153, 20154, 20155, 20156, 20157, 20158, 20963, 20964, 20965, 20966, 20967, 20968, 20969, 20970, 20972, 21192, 21193, 21194;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2212-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2212-2026Class IOngoingVoluntary: Firm initiated