Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2356-2026Class II

NexxZr T Multi A3.5 W98-16mm, REF: 745451

Sagemax Bioceramics, Inc. / initiated 2025-12-11 / Open, Classified / root cause: manufacturing process

Sagemax Bioceramics, Inc. began a recall of NexxZr T Multi A3.5 W98-16mm, REF: 745451 on . FDA classified it as Class II and gives the reason as "Yttrium-stabilized zirconium oxide discs for fixed all-ceramic dental restorations may have reduced or missing dentin layer and when used clinically in the patient's mouth there is an increased risk of fractures under...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2356-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Sagemax Bioceramics, Inc.
Product
NexxZr T Multi A3.5 W98-16mm, REF: 745451
Product code
EIH Powder, Porcelain
Reason for recall
Yttrium-stabilized zirconium oxide discs for fixed all-ceramic dental restorations may have reduced or missing dentin layer and when used clinically in the patient's mouth there is an increased risk of fractures under functional stress.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 12/11/2025, safety notices were emailed to customers who were asked to do the following: 1) Separate and block the affected device in your warehouse so that they can no longer be used. 2) Contact the recalling firm to arrange return. 3) Complete and return the response form via email to [email protected] Queries regarding the measures described in this Field Safety Notice should be addressed to [email protected]
Quantity in commerce
113
Distribution
Worldwide - US Nationwide distribution in the states of CA, MO, MN, NY, FL, ID and the countries of AT, CZ, DE, EE, EG, FR, GR, ID, IT, KZ, RS, SE, SI, TZ, VN.
Product codes and lots
UDI-DI: 00842271175373, Lot: YBDGLD

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2356-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2356-2026Class IIOngoingVoluntary: Firm initiated