Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2465-2026Class II

ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6...

ConMed Corporation / initiated 2026-05-14 / Open, Classified

ConMed Corporation began a recall of ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6 units per pack. Disposable Endoscopic Trocar and... on . FDA classified it as Class II and gives the reason as "Potential for overpressure alerts". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2465-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ConMed Corporation
Product
ConMed AirSeal 12 mm Access Port and Obturator with Blunt Tip, 120mm Length. Catalog Number: iASB12-120. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.
Product code
HIF Insufflator, Laparoscopic
Reason for recall
Potential for overpressure alerts.
FDA root cause
Process change control
Action
ConMed notified consignees on about 05/14/2026 via emailed and mailed letter titled URGENT: Medical Device Correction. Consignees were made aware of the issue and risk associated with the issue. They were instructed that a replacement AirSeal disposable should be available prior to procedures in order to minimize potential delays. Before initiating insufflation, the user should ensure the tubing is properly connected between the AirSeal tube set and Airseal Access Ports. If the connection cannot be securely tightened and the overpressure alert cannot be resolved, please replace the affected device. Detailed instructions were provided on how to connect the AirSeal tube set and Airseal Access Ports. Consignees were also instructed to review inventory for affected units, place a copy of the provided Instructions for Use, Part No. AS-DIS-IFU, Rev. B with any inventory in possession with an active expiration date, contact all applicable departments and facilities that may have received affected products for awareness, and complete and return the medical device response form. If further information relevant to this issue becomes necessary CONMED will communicate as such. On about 06/04/2026 ConMed expanded their recall to include internationally distributed units. The letter sent included the same instructions.
Quantity in commerce
891 packs (5,346 eaches)
Distribution
Nationwide distribution. International distribution to Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, CHILE, CHINA, COLOMBIA, Croatia, Denmark, DOMINICAN REPUBLIC, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, HONG KONG, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Luxembourg, MALAYSIA, Malta, Montenegro, Netherlands, NEW ZEALAND, Norway, Oman, Palestine, PANAMA, PHILIPPINES, Poland, Qatar, Romania, Russia, Saudi Arabia, Serbia, SINGAPORE, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, TAIWAN,PROVINCE OF CHINA, THAILAND, Turkey, Ukraine, UNITED ARAB EMIRATES, United Kingdom. .
Product codes and lots
Catalog Number: iASB12-120. UDI-DI: 10845854045725. Lot Numbers: 202507314, 202506194, 202510024, 202501165, 202504034, 202408225, 202411264. *OUS added 06/10/2026 expansion: 202510024, 202507314, 202504034, 202501165, 202411264, 202408225, 202506194. Manufactured from 14 June 2024 to 12 January 2026.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2465-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2464-2026Class IIOngoingVoluntary: Firm initiated
Z-2465-2026Class IIOngoingVoluntary: Firm initiated
Z-2466-2026Class IIOngoingVoluntary: Firm initiated
Z-2467-2026Class IIOngoingVoluntary: Firm initiated
Z-2468-2026Class IIOngoingVoluntary: Firm initiated
Z-2469-2026Class IIOngoingVoluntary: Firm initiated
Z-2470-2026Class IIOngoingVoluntary: Firm initiated
Z-2471-2026Class IIOngoingVoluntary: Firm initiated
Z-2472-2026Class IIOngoingVoluntary: Firm initiated
Z-2473-2026Class IIOngoingVoluntary: Firm initiated
Z-2474-2026Class IIOngoingVoluntary: Firm initiated