Z-2272-2026Class II

Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044

Medtronic, Inc. / initiated 2026-04-23 / Open, Classified

Covidien LLC began a recall of Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044 on . FDA classified it as Class II and gives the reason as "Temperature probe devices lack FDA clearance". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2272-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044
Product code
BZT Stethoscope, Esophageal, With Electrical Conductors
Reason for recall
Temperature probe devices lack FDA clearance.
FDA root cause
Under Investigation by firm
Action
On April 23, 2026 Covidien (subsidiary of Medtronic) issued a Urgent Medical Device Recall Notification to affected consignees via USPS. Covidien ask consignees to take the following actions: 1. Quarantine all unused product from the affected lots. 2. Return all unused product(s) from the affected lots in your inventory to Medtronic as described in the Customer Confirmation Form. 3. Share this notice with all relevant personnel within your organization, and with any organization to which the potentially affected product from the specified lots has been transferred or distributed. 4. Please complete and return the enclosed Customer Confirmation Form to [email protected] even if you do not have unused inventory.
Quantity in commerce
1775 units
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IL, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, PR, TN, TX, UT, VA, WA, WV.
Product codes and lots
UDI: 30884521828388/ Lot: 25F0945JZX 25H0949JZX, 25I1065JZX, 25I1066JZX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2272-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2271-2026Class IIOngoingVoluntary: Firm initiated
Z-2272-2026Class IIOngoingVoluntary: Firm initiated
Z-2273-2026Class IIOngoingVoluntary: Firm initiated