Z-2311-2026Class II
Plum Solo Precision IV Pump, 40001-0401
Abbott Medical / initiated 2026-04-29 / Open, Classified / root cause: software
ICU Medical, Inc. began a recall of Plum Solo Precision IV Pump, 40001-0401 on . FDA classified it as Class II and gives the reason as "Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2311-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Plum Solo Precision IV Pump, 40001-0401
- Product code
- FRN Pump, Infusion
- Reason for recall
- Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- On April 29, 2026, the firm notified consignees via letters titled "URGEND MEDICAL DEVICE CORRECTION." Customers were instructed to identify all affected pumps in their possession and ensure all users or potential users of the pumps are immediately made aware of the notification. During bolus programming, clinicians should check whether the soft limit override icon is displayed next to the bolus dose entry field on the programming screen. If the pump displays the icon, the clinician should review the program and make any necessary adjustments before confirming the program. The firm is developing a software update to correct the issue and will notify customers when it is available.
- Quantity in commerce
- 13,613
- Distribution
- Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
- Product codes and lots
- UDI-DI M335400011 Plum Solo Precision IV Pump v1.0.3
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2311-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2312-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2313-2026 | Class II | Ongoing | Voluntary: Firm initiated |