Z-2301-2026Class II

BD Pyxis Mini Main REF: 349 Medication cabinet

CareFusion 303, Inc / initiated 2026-04-30 / Open, Classified / root cause: design

CareFusion 303, Inc. began a recall of BD Pyxis Mini Main REF: 349 Medication cabinet on . FDA classified it as Class II and gives the reason as "Fingerprint scanner may overheat to a temperature to cause 1st degree burn". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2301-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis Mini Main REF: 349 Medication cabinet
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
Root cause tag
design
FDA root cause
Component design/selection
Action
On April 30, 2026, CareFusion (subsidiary of Becton, Dickinson and Company) issued a Urgent Medical Device Recall Notification, via mail or Email. CareFusion ask consignees to take the following actions: 1. Dispensing cabinets can continue to be used as intended. 2. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of Scanner and file a complaint. 3. Review and update facility policies and procedures related to situations where the Bio-ID Scanner is unavailable to mitigate delays in care. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form.
Quantity in commerce
709 units
Distribution
US: Nationwide OUS: AE, AU, BE, BH, BR, CA, CL, ES, GB, IT, KW, MA, MX, NZ, PT, QA, SA, SG, TH, TW
Product codes and lots
All Serial Numbers/UDI:10885403512582

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2301-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2297-2026Class IIOngoingVoluntary: Firm initiated
Z-2298-2026Class IIOngoingVoluntary: Firm initiated
Z-2299-2026Class IIOngoingVoluntary: Firm initiated
Z-2300-2026Class IIOngoingVoluntary: Firm initiated
Z-2301-2026Class IIOngoingVoluntary: Firm initiated
Z-2302-2026Class IIOngoingVoluntary: Firm initiated
Z-2303-2026Class IIOngoingVoluntary: Firm initiated
Z-2304-2026Class IIOngoingVoluntary: Firm initiated