Z-2207-2026Class I
Atlan A350. Model Number: 8211500. anesthesia workstation
Draeger, Inc. / initiated 2026-05-07 / Open, Classified
Draeger, Inc. began a recall of Atlan A350. Model Number: 8211500. anesthesia workstation on . FDA classified it as Class I and gives the reason as "The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2207-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Draeger, Inc.
- Product
- Atlan A350. Model Number: 8211500. anesthesia workstation
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- FDA root cause
- Process change control
- Action
- Draeger notified consignees of the recall update and expansion on about 05/07/2026 via emailed letter titled Urgent Medical Device Correction Update. Consignees were made aware that due to continuous monitoring, Draeger has determined that an additional number of Atlan devices are affected by the issue and they are replacing ventilator motors. Consignees were instructed that a Draeger Service Representative will contact consignees to arrange a date for the ventilator motor assembly replacement, and until the corrective action is implemented, the device may continue to be used with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Consignees were also instructed to complete and return the provided Medical Device Correction Return Response Acknowledgment and Receipt Form and if products have been made available to third parties, please forward this Urgent Medical Device Correction notice to them.
- Quantity in commerce
- 3,515 units
- Distribution
- US distribution in WI, AR, MN, WI, CO, MA, PA, AZ, FL, NY, SD, ND. International distribution to: Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chili, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Gabon, Germany, Ghana, Greece, Greenland, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Italy, Israel, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mauritius, Mexico, Morocco, Mozambique, Namibia, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Venezuela, Vietnam, Yemen
- Product codes and lots
- Model Number: 8211500. UDI-DI Number: 04048675556176.Serial Numbers: ASSL-0347, ASSL-0348, ASSL-0351, ASSH-0220, ASSH-0221, ASSH-0224, ASSH-0225, ASSH-0228, ASSH-0229, ASSH-0230, ASSH-0234, ASSH-0235, ASTA-0086, ASSL-0349, ASSL-0350, ASSL-0352, ASSL-0353, ASSH-0222, ASSH-0223, ASSH-0226, ASSH-0227, ASSH-0231, ASSH-0232, ASSH-0233, ASSH-0236, ASSH-0237, ASSH-0238, ASSN-0185, ASSN-0186, ASSN-0187, ASSN-0188, ASSN-0189, ASSN-0190, ASTB-0037, ASTB-0040, ASTA-0087, ASTA-0088, ASSM-0130, ASSM-0131, ASSM-0132.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2207-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2208-2026 | Class I | Ongoing | Voluntary: Firm initiated |