Z-2243-2026Class II

GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF...

GE Medical Systems Information Technologies, Inc. / initiated 2026-04-24 / Open, Classified / root cause: software

GE Medical Systems Information Technologies Inc began a recall of GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server... on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2243-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Medical Systems Information Technologies, Inc.
Product
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Root cause tag
software
FDA root cause
Software design
Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 4/24/2026 by letter using a traceable means. The notice explained the the safety issue pertaining to the persistent NO COMM or OFF NETWORK condition, and provided instructions for the safe use of the device should either issue arise. "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your ApexPro CTS with Central Station for telemetry only or combination monitoring unless you observe a persistent (more than a few seconds) NO COMM or OFF NETWORK notification. If you do observe a persistent NO COMM or OFF NETWORK notification, follow the instructions below. 1. Use alternative bedside patient monitoring to ensure continuous clinical surveillance until the NO COMM or OFF NETWORK condition is resolved by GE HealthCare. NOTE: If you are using combination monitoring, it is recommended to switch ECG monitoring to the bedside monitor. 2. Immediately notify your Facility s Biomedical Engineering, IT or Service Department and contact GE HealthCare Service. 3. Verify monitoring is restored at the Central Station(s) and/or bedside monitor(s). Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please post this notification by your Central Station(s). Please retain this document for your records." "GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
14 units
Distribution
Worldwide distribution.
Product codes and lots
UDI none, Serial number or Sales Order number SS921120014SA, SS921120017SA, SS921120021SA, SS921120018SA, SS921120020SA, SS920390001SA, SS921120007SA, SS921120008SA, SS921120012SA, SS921120016SA, SS920500001SA, SS920500002SA, SS921120009SA, SS921120010SA.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2243-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2242-2026Class IIOngoingVoluntary: Firm initiated
Z-2243-2026Class IIOngoingVoluntary: Firm initiated
Z-2244-2026Class IIOngoingVoluntary: Firm initiated
Z-2245-2026Class IIOngoingVoluntary: Firm initiated
Z-2246-2026Class IIOngoingVoluntary: Firm initiated
Z-2247-2026Class IIOngoingVoluntary: Firm initiated