Z-2284-2026Class II

Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and...

Philips Medical Systems Nederland B.V. / initiated 2023-12-07 / Open, Classified / root cause: software

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key on . FDA classified it as Class II and gives the reason as "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2284-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Root cause tag
software
FDA root cause
Software design
Action
Philips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to: - Keep this Important Product Notice with the documentation of the system until Philips corrects your system. - Circulate this notice to all users of the system so that they are aware of the issue. - Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Important Product Notice. For further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
Quantity in commerce
16 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI (01)00884838099203(21)3, (01)00884838099203(21)7, (01)00884838085275(21)131, (01)00884838099203(21)54, (01)00884838085275(21)84, (01)00884838099203(21)6, (01)00884838099203(21)93, (01)00884838085275(21)132, (01)00884838099203(21)69, (01)00884838099203(21)9, (01)00884838099203(21)115, (01)00884838099203(21)85, (01)00884838099203(21)19, (01)00884838099203(21)49, (01)00884838099203(21)108, (01)00884838085275(21)108.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2284-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2284-2026Class IIOngoing
Z-2285-2026Class IIOngoing
Z-2286-2026Class IIOngoing
Z-2287-2026Class IIOngoing
Z-2288-2026Class IIOngoing
Z-2289-2026Class IIOngoing
Z-2290-2026Class IIOngoing
Z-2291-2026Class IIOngoing