Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5...
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key on . FDA classified it as Class II and gives the reason as "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2288-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to: - Keep this Important Product Notice with the documentation of the system until Philips corrects your system. - Circulate this notice to all users of the system so that they are aware of the issue. - Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Important Product Notice. For further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
- Quantity in commerce
- 51 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI (01)00884838099265(21)104, (01)00884838085350(21)221, (01)00884838099265(21)153, (01)00884838099265(21)46, (01)00884838099265(21)28, (01)00884838085350(21)250, (01)00884838099265(21)72, (01)00884838099265(21)208, (01)00884838099265(21)41, (01)00884838099265(21)122, (01)00884838099265(21)147, (01)00884838085350(21)225, (01)00884838099265(21)262, (01)00884838099265(21)112, (01)00884838099265(21)110, (01)00884838099265(21)207, (01)00884838099265(21)68, (01)00884838099265(21)177, (01)00884838099265(21)29, (01)00884838099265(21)71, (01)00884838099265(21)37, (01)00884838099265(21)175, (01)00884838099265(21)148, (01)00884838099265(21)205 (01)00884838099265(21)229, (01)00884838099265(21)228, (01)00884838085350(21)277, (01)00884838085350(21)115, (01)00884838099265(21)157, (01)00884838085350(21)163, (01)00884838099265(21)268, (01)00884838085350(21)75, (01)00884838085350(21)132, (01)00884838085350(21)147, (01)00884838085350(21)208, (01)00884838085350(21)104, (01)00884838085350(21)244, (01)00884838085350(21)234, (01)00884838099265(21)53, (01)00884838099265(21)259, (01)00884838099265(21)55, (01)00884838099265(21)235, (01)00884838099265(21)298, (01)00884838085350(21)70, (01)00884838099265(21)233, (01)00884838099265(21)141, (01)00884838099265(21)289, (01)00884838099265(21)249, (01)00884838099265(21)84, (01)00884838099265(21)111, (01)00884838099265(21)227.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2284-2026 | Class II | Ongoing | |||
| Z-2285-2026 | Class II | Ongoing | |||
| Z-2286-2026 | Class II | Ongoing | |||
| Z-2287-2026 | Class II | Ongoing | |||
| Z-2288-2026 | Class II | Ongoing | |||
| Z-2289-2026 | Class II | Ongoing | |||
| Z-2290-2026 | Class II | Ongoing | |||
| Z-2291-2026 | Class II | Ongoing |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.