Z-2416-2026Class II
Block Tray Dynjra2119a Indwelling Nerve Block Dynjra1011a
MEDLINE INDUSTRIES, LP - Northfield / initiated 2026-04-27 / Open, Classified / root cause: manufacturing process
Medline Industries, LP began a recall of BLOCK TRAY DYNJRA2119A INDWELLING NERVE BLOCK DYNJRA1011A on . FDA classified it as Class II and gives the reason as "Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2416-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Block Tray Dynjra2119a Indwelling Nerve Block Dynjra1011a
- Product code
- OGI Nasal Endotracheal Tube Holder Kit
- Reason for recall
- Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity in commerce
- 49,654 kits total
- Distribution
- US Nationwide distribution. OUS distribution pending.
- Product codes and lots
- DYNJRA2119A UDI-DI 10198459194702 lot 25LBB738 DYNJRA1011A UDI-DI 10888277783614 lots 25KBB849 25LBB734
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2414-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2415-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2416-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2417-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2418-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2419-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2420-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2421-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2422-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2423-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2424-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2425-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2426-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2427-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2428-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2429-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2430-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2431-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2432-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2433-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2434-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2435-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2436-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2437-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2438-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2439-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2440-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2441-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2442-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2443-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2444-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2445-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2446-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2447-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.