Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2417-2026Class II

Minor Cath Lab Pack Dynj84775a or Vascular Angio Pack Dynj47269d Port-a-Cath Dynj63276g

MEDLINE INDUSTRIES, LP - Northfield / initiated 2026-04-27 / Open, Classified / root cause: manufacturing process

Medline Industries, LP began a recall of MINOR CATH LAB PACK DYNJ84775A OR VASCULAR ANGIO PACK DYNJ47269D PORT-A-CATH DYNJ63276G on . FDA classified it as Class II and gives the reason as "Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2417-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Minor Cath Lab Pack Dynj84775a or Vascular Angio Pack Dynj47269d Port-a-Cath Dynj63276g
Product code
OES Cardiac Catheterization Kit
Reason for recall
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in commerce
49,654 kits total
Distribution
US Nationwide distribution. OUS distribution pending.
Product codes and lots
DYNJ84775A UDI-DI 10195327606527 lot 25LMH070 DYNJ47269D UDI-DI 10195327674786 lot 25LMJ234 DYNJ63276G UDI-DI 10198459458835 lot 26BBC598

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2417-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2414-2026Class IIOngoingVoluntary: Firm initiated
Z-2415-2026Class IIOngoingVoluntary: Firm initiated
Z-2416-2026Class IIOngoingVoluntary: Firm initiated
Z-2417-2026Class IIOngoingVoluntary: Firm initiated
Z-2418-2026Class IIOngoingVoluntary: Firm initiated
Z-2419-2026Class IIOngoingVoluntary: Firm initiated
Z-2420-2026Class IIOngoingVoluntary: Firm initiated
Z-2421-2026Class IIOngoingVoluntary: Firm initiated
Z-2422-2026Class IIOngoingVoluntary: Firm initiated
Z-2423-2026Class IIOngoingVoluntary: Firm initiated
Z-2424-2026Class IIOngoingVoluntary: Firm initiated
Z-2425-2026Class IIOngoingVoluntary: Firm initiated
Z-2426-2026Class IIOngoingVoluntary: Firm initiated
Z-2427-2026Class IIOngoingVoluntary: Firm initiated
Z-2428-2026Class IIOngoingVoluntary: Firm initiated
Z-2429-2026Class IIOngoingVoluntary: Firm initiated
Z-2430-2026Class IIOngoingVoluntary: Firm initiated
Z-2431-2026Class IIOngoingVoluntary: Firm initiated
Z-2432-2026Class IIOngoingVoluntary: Firm initiated
Z-2433-2026Class IIOngoingVoluntary: Firm initiated
Z-2434-2026Class IIOngoingVoluntary: Firm initiated
Z-2435-2026Class IIOngoingVoluntary: Firm initiated
Z-2436-2026Class IIOngoingVoluntary: Firm initiated
Z-2437-2026Class IIOngoingVoluntary: Firm initiated
Z-2438-2026Class IIOngoingVoluntary: Firm initiated
Z-2439-2026Class IIOngoingVoluntary: Firm initiated
Z-2440-2026Class IIOngoingVoluntary: Firm initiated
Z-2441-2026Class IIOngoingVoluntary: Firm initiated
Z-2442-2026Class IIOngoingVoluntary: Firm initiated
Z-2443-2026Class IIOngoingVoluntary: Firm initiated
Z-2444-2026Class IIOngoingVoluntary: Firm initiated
Z-2445-2026Class IIOngoingVoluntary: Firm initiated
Z-2446-2026Class IIOngoingVoluntary: Firm initiated
Z-2447-2026Class IIOngoingVoluntary: Firm initiated