Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2504-2026Class II

MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2026-05-04 / Open, Classified / root cause: manufacturing process

Medline Industries, LP began a recall of MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C on . FDA classified it as Class II and gives the reason as "During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2504-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.
Product code
OHM Cesarean Section Tray
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Quantity in commerce
168 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Product codes and lots
Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21KME435; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMC247; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMB678; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21IMH919; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24CBO389; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24ABR872.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2504-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2479-2026Class IIOngoingVoluntary: Firm initiated
Z-2480-2026Class IIOngoingVoluntary: Firm initiated
Z-2481-2026Class IIOngoingVoluntary: Firm initiated
Z-2482-2026Class IIOngoingVoluntary: Firm initiated
Z-2483-2026Class IIOngoingVoluntary: Firm initiated
Z-2484-2026Class IIOngoingVoluntary: Firm initiated
Z-2485-2026Class IIOngoingVoluntary: Firm initiated
Z-2486-2026Class IIOngoingVoluntary: Firm initiated
Z-2487-2026Class IIOngoingVoluntary: Firm initiated
Z-2488-2026Class IIOngoingVoluntary: Firm initiated
Z-2489-2026Class IIOngoingVoluntary: Firm initiated
Z-2490-2026Class IIOngoingVoluntary: Firm initiated
Z-2491-2026Class IIOngoingVoluntary: Firm initiated
Z-2492-2026Class IIOngoingVoluntary: Firm initiated
Z-2493-2026Class IIOngoingVoluntary: Firm initiated
Z-2494-2026Class IIOngoingVoluntary: Firm initiated
Z-2495-2026Class IIOngoingVoluntary: Firm initiated
Z-2496-2026Class IIOngoingVoluntary: Firm initiated
Z-2497-2026Class IIOngoingVoluntary: Firm initiated
Z-2498-2026Class IIOngoingVoluntary: Firm initiated
Z-2499-2026Class IIOngoingVoluntary: Firm initiated
Z-2500-2026Class IIOngoingVoluntary: Firm initiated
Z-2501-2026Class IIOngoingVoluntary: Firm initiated
Z-2502-2026Class IIOngoingVoluntary: Firm initiated
Z-2503-2026Class IIOngoingVoluntary: Firm initiated
Z-2504-2026Class IIOngoingVoluntary: Firm initiated
Z-2505-2026Class IIOngoingVoluntary: Firm initiated
Z-2506-2026Class IIOngoingVoluntary: Firm initiated
Z-2507-2026Class IIOngoingVoluntary: Firm initiated
Z-2508-2026Class IIOngoingVoluntary: Firm initiated