Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2913-2026Class II

navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5...

Ventana Medical Systems Inc / initiated 2026-04-20 / Open, Classified / root cause: software

Ventana Medical Systems, Inc. began a recall of navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code... on . FDA classified it as Class II and gives the reason as "Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2913-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ventana Medical Systems Inc
Product
navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code: 761333602095AQ In the United States, navify Digital Pathology software (previously known as uPath Enterprise Software) is a component of the Roche Digital Pathology Dx device cleared by the US FDA for Whole Slide Imaging purposes. Roche Digital Pathology Dx is an automated digital slide creation, viewing, and management system.
Product code
PSY Whole Slide Imaging System
Reason for recall
Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window
Root cause tag
software
FDA root cause
Software design
Action
On 05/01/2026, the firm sent via UPS Ground an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that Roche eceived two complaints outside of the U.S. concerning the navify Digital Pathology (nDP) image management system regarding an issue where images may be mismatched or associated with incorrect specimen and slide labels within a given case. The discrepancy occurs exclusively within a subset of individual cases in which a specific nomenclature is used for specimen names and the case contains multiple specimens. Customers are instructed to: " Until navify Digital Pathology software version 2.5.2 goes live on May 8th at 6:30 PM PDT (9:30 PM EDT), users should zoom into or select slide images and cross-reference specimen/block/slide labels with the correct, synchronized information in the top Case Bar, Thumbnail Panel, or Slide Info Panel in the Viewer window before proceeding with clinical analysis. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " Complete all sections of the enclosed recall response form (TP-02782) and fax it to 1-833-237-1015 or email it to [email protected]. " Distribute this UMDC within your facility and network to ensure relevant personnel are aware of the issue. " File this Urgent Medical Device Correction (UMDC) for future reference. For Questions - contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336
Quantity in commerce
190
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, WI, and WV. The countries of Austria, Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Romania, Spain, and Switzerland, Argentina, Bosnia & Herzegovina, Brazil, Chile, Colombia, Dominican Republic, Ecuador, Jordan, Mexico, Myanmar, Peru, Turkey, United Kingdom, Australia, Bahrain, Colombia, Hong Kong, Portugal and India.
Product codes and lots
navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code: 761333602095AQ Software version Numbers: 2.5.0, 2.5.0.1, 2.5.0.2, and 2.5.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2913-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2913-2026Class IIOngoingVoluntary: Firm initiated