Medline convenience kits: BASIC EYE DYNJ905154I TUBAL LIGATION CDS CDS760053K
Medline Industries, LP began a recall of Medline convenience kits: BASIC EYE DYNJ905154I TUBAL LIGATION CDS CDS760053K on . FDA classified it as Class II and gives the reason as "Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2435-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline convenience kits: BASIC EYE DYNJ905154I TUBAL LIGATION CDS CDS760053K
- Product code
- OJK Eye Tray
- Reason for recall
- Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity in commerce
- 49,654 kits total
- Distribution
- US Nationwide distribution. OUS distribution pending.
- Product codes and lots
- DYNJ905154I UDI-DI 10198459549977 lots 26ABB308 26BBA671 CDS760053K UDI-DI 10195327182977 lot 25KBP010
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2414-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2415-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2416-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2417-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2418-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2419-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2420-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2421-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2422-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2423-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2424-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2425-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2426-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2427-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2428-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2429-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2430-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2431-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2432-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2433-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2434-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2435-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2436-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2437-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2438-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2439-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2440-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2441-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2442-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2443-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2444-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2445-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2446-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2447-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.