Z-2600-2026Class I
Abiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set
Abiomed, Inc. / initiated 2026-05-22 / Open, Classified
Abiomed, Inc. began a recall of Abiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000435. Introducer Product Code: 1000435 on . FDA classified it as Class I and gives the reason as "Potential for thrombus formation during prolonged use of the introducer". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2600-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abiomed, Inc.
- Product
- Abiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product Code: 1000435. Introducer Product Code: 1000435.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Potential for thrombus formation during prolonged use of the introducer.
- FDA root cause
- Under Investigation by firm
- Action
- Abiomed notified consignees on about 05/22/2026, via letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were made aware of the issue and that if thrombus formation were to occur, the user may encounter difficulty aspirating the sheath sidearm or valve and may observe thrombus in the syringe or sidearm post aspiration or around the valve post CP pump removal. They were instructed to review inventory for affected units and quarantine them, arrange for their return using the provided shipment label from Sedgwick, and complete and return the provided business response form. Consignees were also instructed to forward the notification to anyone in the facility that needs to be informed, if any of the subject products have been forwarded to another facility, contact that facility and provide them with the notification, and to post a copy of the notification in a visible area for awareness.
- Quantity in commerce
- 17 units
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
- Product codes and lots
- Impella Set Product Code: 1000435; GTIN: 00813502013009; Serial Numbers: N/A. Introducer Product Code: 1000435; Batch Numbers: 2038707, 2045782.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2599-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2600-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2601-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2602-2026 | Class I | Ongoing | Voluntary: Firm initiated |