Z-2601-2026Class I
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer...
Abiomed, Inc. / initiated 2026-05-22 / Open, Classified
Abiomed, Inc. began a recall of Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834 on . FDA classified it as Class I and gives the reason as "Potential for thrombus formation during prolonged use of the introducer". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2601-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abiomed, Inc.
- Product
- Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Potential for thrombus formation during prolonged use of the introducer.
- FDA root cause
- Under Investigation by firm
- Action
- Abiomed notified consignees on about 05/22/2026, via letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were made aware of the issue and that if thrombus formation were to occur, the user may encounter difficulty aspirating the sheath sidearm or valve and may observe thrombus in the syringe or sidearm post aspiration or around the valve post CP pump removal. They were instructed to review inventory for affected units and quarantine them, arrange for their return using the provided shipment label from Sedgwick, and complete and return the provided business response form. Consignees were also instructed to forward the notification to anyone in the facility that needs to be informed, if any of the subject products have been forwarded to another facility, contact that facility and provide them with the notification, and to post a copy of the notification in a visible area for awareness.
- Quantity in commerce
- 168 units
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
- Product codes and lots
- 1. Impella Set Product Code: 1000413; GTIN: 00813502013566; Serial numbers: 649727, 649728, 649729, 649730, 649731, 654349, 654348, 654342, 654341, 654340, 654339, 654338, 654336, 654335, 654334, 654333, 653010, 653009, 653008, 653007, 653006, 653005, 653004, 653003, 653002, 653001, 655735, 655733, 654357, 654356, 654382, 654383, 654381, 654380, 654378, 654367, 654366, 654364, 654365, 654362, 654363, 654361, 654360, 654359, 654358, 654379, 655734, 654386, 654355, 654385, 654384, 654353, 654352, 654351, 654350, 654387, 657649, 657648, 657646, 657647, 657645, 657644, 655742, 655739, 655741, 655740, 655738, 655737, 655736, 656743, 656738, 653050, 653049, 653048, 653047, 653045, 653044, 653043, 653042, 653041. Introducer Product Code: 2000684; Batch Numbers: 2026724558, 2027818724. 2. Product Code: 1000834. GTIN: 00813502013948. Serial numbers: 653474, 653471, 653485, 653470, 653469, 653468, 653467, 653465, 653466, 653463, 653464, 653482, 653483, 653484, 653481, 653490, 653462, 653478, 653479, 653489, 653488, 653487, 653486, 653476, 653477, 656002, 656001, 656000, 655998, 655997, 655993, 655996, 655995, 655994, 656737, 656736, 656735, 656734, 656733, 656730, 656731, 656729, 656728, 656346, 656343, 656342, 656341, 656340, 656338, 659804, 659795, 656347, 650264, 659801, 659802, 659800, 659799, 659798, 656363, 656367, 656362, 656361, 656360, 656359, 656358, 659794, 659793, 659791, 659792, 659790, 659789, 659788, 659786, 659787, 659785, 661408, 661407, 661405, 661404, 661392, 661391, 661389, 661403, 661402, 661401, 661400, 662657, 662654. Introducer Product Code: 2000785; Batch Numbers: 2026787668, 2027818660. Distributed in Japan.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2599-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2600-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2601-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2602-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
