Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6...
United Orthopedic Corp. began a recall of Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and... on . FDA classified it as Class II and gives the reason as "Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2884-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- United Orthopedic Corp.
- Product
- Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and 9325-5721 thru 9325-5726 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A
- Product code
- KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
- Reason for recall
- Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On 03/11/2026, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due the fact that the PSA Box Trial cannot be properly assembled with the Hinge Trial in the affected devices because of incomplete processing of the screw components, which may result in interference during assembly with the corresponding device. On 05/20/2026, the firm emailed an updated "URGENT: MEDCIAL DEVICE RECALL" Letter Customers are instructed: 1. Examine your inventory immediately to locate the affected product whose quantities and lot numbers are listed above. Please note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in your territory for immediate return. 3. Customer Service will be contacting you regarding replacements. If you still have the product, you are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if your firm has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For Question - contact UOC Sr. Director of Sales Operations at 949.328.3366.
- Quantity in commerce
- 2
- Distribution
- US Nationwide distribution in the states of Texas and Michigan.
- Product codes and lots
- Product Number; UDI-DI Code; Lot Code: Product number 9325-5711; UDI-DI 04719886907032; Lot number 25FB41 Product number 9325-5712; UDI-DI 04719886907049; Lot number 25FB42 Product number 9325-5713; UDI-DI 04719886907056; Lot number 25FB43 Product number 9325-5714; UDI-DI 04719886907063; Lot number 25FB44 Product number 9325-5715; UDI-DI 04719886907070; Lot number 25FB45 Product number 9325-5716; UDI-DI 04719886907087; Lot number 25FB46 Product number 9325-5721; UDI-DI 04719886907094; Lot number 25FB35 Product number 9325-5722; UDI-DI 04719886907100; Lot number 25FB36 Product number 9325-5723; UDI-DI 04719886907117; Lot number 25FB37 Product number 9325-5724; UDI-DI 04719886907124; Lot number 25FB38 Product number 9325-5725; UDI-DI 04719886907131; Lot number 25FB39 Product number 9325-5726; UDI-DI 04719886907148; Lot number 25FB40
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2883-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2884-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2885-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2886-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.