Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2451-2026Class II

Medtronic O-arm O2 Imaging System. Model Number: BI70002000

Medtronic Navigation, Inc / initiated 2026-05-19 / Open, Classified / root cause: software

Medtronic Navigation, Inc.-Boxborough began a recall of Medtronic O-arm O2 Imaging System. Model Number: BI70002000 on . FDA classified it as Class II and gives the reason as "Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2451-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic O-arm O2 Imaging System. Model Number: BI70002000.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Medtronic notified consignees on about 05/19/2026 via "URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION" letter. Consignees were informed that affected systems may continue to be used prior to service by Medtronic to update the firmware. The issue is intermittent and may not present, however if it does it can be resolved by power cycling (rebooting) the system. A Medtronic Field Service Engineer will contact each facility to schedule time to update the firmware on each system to permanently resolve this issue. Consignees were instructed to complete and return the provided Customer Confirmation Form, provide the notification to all applicable personnel within the organization or to customers if product has been transferred, and to maintain a copy of the notice with their records.
Quantity in commerce
589 units
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Libya, Malaysia, Mexico, Morocco, Nepal, Netherlands, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Product codes and lots
Model Number: BI70002000. GTIN (Serial Numbers): 00643169639683 (C1640), 00643169651814 (C3815, C3816, C3817, C3818, C3891, C3898, C3899, C3900, C3901, C3902, C3903, C3905, C3990, C3997, C3998, C3999, C4001, C4002, C4003, C4004, C4059, C4060, C4061, C4062, C4063, C4064, C4065, C4066, C4067, C4068, C4116, C4117, C4118, C4119, C4120, C4121, C4122, C4123, C4130, C4201, C4202, C4203, C4356, C4357, C4360, C4361), 00763000132378 (C2096, C2187), 00763000355555 (C2768), 00763000481599 (C3895, C3996), 00763000496586 (C3794, C3896, C4023, C4031, C4032, C4043, C4088, C4092), 00763000496593 (C3887, C3889, C3978, C3988, C3991, C3994, C4041, C4109, C4110, C4111, C4161), 00763000496609 (C3802), 00763000496630 (C4035), 00763000496647 (C4012, C4126, C4129), 00763000496661 (C3824, C3987, C4071, C4072), 00763000496678 (C3879), 00763000496692 (C3799, C3823, C3897, C3979, C4027, C4037, C4044, C4125), 00763000496708 (C3825), 00763000496715 (C4074, C4128), 00763000496722 (C3812, C3822, C3828, C3881, C4034, C4036, C4038, C4039, C4160), 00763000616304 (C3795), 00763000616397 (C3800), 00763000616441 (C3811), 00763000616458 (C3783, C3784), 00763000616564 (C3316, C3735, C3777, C3778, C3779, C3780, C3781, C3786, C3787, C3790, C3791, C3792, C3793, C3796, C3797, C3798), 00763000709365 (C4307), 00763000709372 (C4221, C4284, C4294, C4301, C4354, C4370, C4399), 00763000709389 (C4166, C4295, C4402), 00763000709396 (C4394), 00763000709402 (C4149, C4150), 00763000709419 (C4395), 00763000709426 (C4115, C4248, C4304), 00763000709457 (C4293, C4298,C4403), 00763000709464 (C4368), 00763000709488 (C4204, C4205, C4302, C4363, C4366), 00763000709518 (C4148, C4167, C4222, C4223, C4224, C4225, C4285, C4292, C4303, C4309, C4362, C4369, C4374, C4375), 00763000709532 (C4206, C4208, C4209, C4236, C4245, C4283, C4286, C4287, C4288, C4289, C4290, C4308, C4310, C4372, C4373, C4377, C4378), 00763000709556 (C4390), 00763000709570 (C4249, C4250), 00763000709594 (C4396), 00763000709631 (C3883, C3884, C3885, C3907, C3908, C3909, C3910, C3911, C3912, C3913, C3914, C3915, C3916, C3918, C3919, C3920, C3921, C3922, C3923, C3924, C3925, C3926, C3927, C3928, C3929, C3930, C3931, C3932, C3933, C3934, C3935, C3936, C3937, C3938, C3939, C3940, C3941, C3942, C3943, C3944, C3945, C3946, C3947, C3948, C3949, C3950, C3951, C3953, C3954, C3955, C3956, C3957, C3958, C3959, C3960, C3961, C3962, C3963, C3964, C3965, C3966, C3967, C3968, C3969, C3970, C3971, C3972, C3973, C3974, C3975, C3976, C3977, C3982, C3983, C3984, C4014, C4015, C4016, C4017, C4018, C4019, C4020, C4021, C4022, C4028, C4045, C4046, C4047, C4048, C4049, C4050, C4051, C4052, C4053, C4075, C4076, C4077, C4078, C4079, C4080, C4081, C4082, C4083, C4084, C4085, C4086, C4087, C4089, C4090, C4093, C4094, C4095, C4096, C4097, C4098, C4100, C4101, C4102, C4103, C4104, C4105, C4106, C4107, C4108, C4131, C4132, C4133, C4134, C4135, C4136, C4137, C4138, C4139, C4140, C4141, C4142, C4143, C4144, C4145, C4146, C4151, C4152, C4153, C4154, C4155, C4156, C4157, C4158, C4162, C4163, C4168, C4169, C4170, C4171, C4172, C4173, C4174, C4175, C4176, C4177, C4178, C4179, C4180, C4181, C4182, C4183, C4184, C4185, C4186, C4187, C4188, C4190, C4191, C4193, C4194, C4195, C4196, C4197, C4198, C4199, C4200, C4210, C4211, C4212, C4213, C4214, C4215, C4217, C4218, C4219, C4220, C4226, C4227, C4228, C4229, C4230, C4231, C4232, C4233, C4234, C4235, C4237, C4238, C4239, C4240, C4241, C4242, C4243, C4244, C4246, C4247, C4251, C4252, C4253, C4254, C4256, C4257, C4258, C4259, C4260, C4261, C4263, C4264, C4265, C4266, C4267, C4268, C4269, C4270, C4271, C4272, C4273, C4274, C4278, C4279, C4281, C4311, C4312, C4313, C4314, C4315, C4316, C4318, C4319, C4320, C4321, C4322, C4323, C4324, C4325, C4326, C4328, C4329, C4330, C4331, C4332, C4333, C4334, C4335, C4336, C4337, C4339, C4340, C4341, C4342, C4343, C4344, C4345, C4346, C4347, C4348, C4349, C4350, C4351, C4382, C4383, C4384, C4385, C4386, C4388, C4405, C4406, C4407, C4414, C4415, C4416, C4418, C4421, C4422, C4280), 00763000709648 (C4282, C4300), 00763000709662 (C4306, C4353), 00763000709686 (C4352), 00763000709693 (C4299, C4397), 00763000745462 (C4040), 00763000745486 (C3863, C4069, C4070), 00763000863203 (C4033), 00763000863210 (C3789, C3801, C3804, C3805, C3806, C3807, C3808, C3809, C3878, C3985, C4005, C4006, C4007, C4008, C4009, C4010, C4011), 00763000863241 (C3833, C3890, C3892, C3894, C3980, C3995, C4013, C4029), 00763000863265 (C3826, C4164, C4165), 00763000863272 (C3813, C3814, C3820, C3821, C3827, C3888, C3904, C3906, C3992, C4024, C4026, C4054, C4055, C4056, C4057, C4073, C4112, C4113, C4124, C4147, C4159), 00763000863302 (C3829, C3830, C3831, C3832, C3835, C3837, C3838, C3839, C3840, C3841, C3842, C3843, C3844, C3845, C3848, C3849, C3850, C3851, C3852, C3853, C3854, C3855, C3856, C3857, C3858, C3859, C3860, C3861, C3862, C3864, C3865, C3866, C3867, C3868, C3869, C3870, C3871, C3872, C3873, C3874, C3875, C3876), 00763000871109 (C4127), 00763000871116 (C3989), 00763000871123 (C3986, C3993, C4114), 00763000871130 (C3834, C4058), 00763000967314 (C4389), 00763000967321 (C4296), 00763000987701 (C4364, C4365), 00763000987732 (C4216).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2451-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2451-2026Class IIOngoingVoluntary: Firm initiated