Z-2451-2026Class II
Medtronic O-arm O2 Imaging System. Model Number: BI70002000
Medtronic Navigation, Inc / initiated 2026-05-19 / Open, Classified / root cause: software
Medtronic Navigation, Inc.-Boxborough began a recall of Medtronic O-arm O2 Imaging System. Model Number: BI70002000 on . FDA classified it as Class II and gives the reason as "Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2451-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic O-arm O2 Imaging System. Model Number: BI70002000.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- Medtronic notified consignees on about 05/19/2026 via "URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION" letter. Consignees were informed that affected systems may continue to be used prior to service by Medtronic to update the firmware. The issue is intermittent and may not present, however if it does it can be resolved by power cycling (rebooting) the system. A Medtronic Field Service Engineer will contact each facility to schedule time to update the firmware on each system to permanently resolve this issue. Consignees were instructed to complete and return the provided Customer Confirmation Form, provide the notification to all applicable personnel within the organization or to customers if product has been transferred, and to maintain a copy of the notice with their records.
- Quantity in commerce
- 589 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Libya, Malaysia, Mexico, Morocco, Nepal, Netherlands, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Product codes and lots
- Model Number: BI70002000. GTIN (Serial Numbers): 00643169639683 (C1640), 00643169651814 (C3815, C3816, C3817, C3818, C3891, C3898, C3899, C3900, C3901, C3902, C3903, C3905, C3990, C3997, C3998, C3999, C4001, C4002, C4003, C4004, C4059, C4060, C4061, C4062, C4063, C4064, C4065, C4066, C4067, C4068, C4116, C4117, C4118, C4119, C4120, C4121, C4122, C4123, C4130, C4201, C4202, C4203, C4356, C4357, C4360, C4361), 00763000132378 (C2096, C2187), 00763000355555 (C2768), 00763000481599 (C3895, C3996), 00763000496586 (C3794, C3896, C4023, C4031, C4032, C4043, C4088, C4092), 00763000496593 (C3887, C3889, C3978, C3988, C3991, C3994, C4041, C4109, C4110, C4111, C4161), 00763000496609 (C3802), 00763000496630 (C4035), 00763000496647 (C4012, C4126, C4129), 00763000496661 (C3824, C3987, C4071, C4072), 00763000496678 (C3879), 00763000496692 (C3799, C3823, C3897, C3979, C4027, C4037, C4044, C4125), 00763000496708 (C3825), 00763000496715 (C4074, C4128), 00763000496722 (C3812, C3822, C3828, C3881, C4034, C4036, C4038, C4039, C4160), 00763000616304 (C3795), 00763000616397 (C3800), 00763000616441 (C3811), 00763000616458 (C3783, C3784), 00763000616564 (C3316, C3735, C3777, C3778, C3779, C3780, C3781, C3786, C3787, C3790, C3791, C3792, C3793, C3796, C3797, C3798), 00763000709365 (C4307), 00763000709372 (C4221, C4284, C4294, C4301, C4354, C4370, C4399), 00763000709389 (C4166, C4295, C4402), 00763000709396 (C4394), 00763000709402 (C4149, C4150), 00763000709419 (C4395), 00763000709426 (C4115, C4248, C4304), 00763000709457 (C4293, C4298,C4403), 00763000709464 (C4368), 00763000709488 (C4204, C4205, C4302, C4363, C4366), 00763000709518 (C4148, C4167, C4222, C4223, C4224, C4225, C4285, C4292, C4303, C4309, C4362, C4369, C4374, C4375), 00763000709532 (C4206, C4208, C4209, C4236, C4245, C4283, C4286, C4287, C4288, C4289, C4290, C4308, C4310, C4372, C4373, C4377, C4378), 00763000709556 (C4390), 00763000709570 (C4249, C4250), 00763000709594 (C4396), 00763000709631 (C3883, C3884, C3885, C3907, C3908, C3909, C3910, C3911, C3912, C3913, C3914, C3915, C3916, C3918, C3919, C3920, C3921, C3922, C3923, C3924, C3925, C3926, C3927, C3928, C3929, C3930, C3931, C3932, C3933, C3934, C3935, C3936, C3937, C3938, C3939, C3940, C3941, C3942, C3943, C3944, C3945, C3946, C3947, C3948, C3949, C3950, C3951, C3953, C3954, C3955, C3956, C3957, C3958, C3959, C3960, C3961, C3962, C3963, C3964, C3965, C3966, C3967, C3968, C3969, C3970, C3971, C3972, C3973, C3974, C3975, C3976, C3977, C3982, C3983, C3984, C4014, C4015, C4016, C4017, C4018, C4019, C4020, C4021, C4022, C4028, C4045, C4046, C4047, C4048, C4049, C4050, C4051, C4052, C4053, C4075, C4076, C4077, C4078, C4079, C4080, C4081, C4082, C4083, C4084, C4085, C4086, C4087, C4089, C4090, C4093, C4094, C4095, C4096, C4097, C4098, C4100, C4101, C4102, C4103, C4104, C4105, C4106, C4107, C4108, C4131, C4132, C4133, C4134, C4135, C4136, C4137, C4138, C4139, C4140, C4141, C4142, C4143, C4144, C4145, C4146, C4151, C4152, C4153, C4154, C4155, C4156, C4157, C4158, C4162, C4163, C4168, C4169, C4170, C4171, C4172, C4173, C4174, C4175, C4176, C4177, C4178, C4179, C4180, C4181, C4182, C4183, C4184, C4185, C4186, C4187, C4188, C4190, C4191, C4193, C4194, C4195, C4196, C4197, C4198, C4199, C4200, C4210, C4211, C4212, C4213, C4214, C4215, C4217, C4218, C4219, C4220, C4226, C4227, C4228, C4229, C4230, C4231, C4232, C4233, C4234, C4235, C4237, C4238, C4239, C4240, C4241, C4242, C4243, C4244, C4246, C4247, C4251, C4252, C4253, C4254, C4256, C4257, C4258, C4259, C4260, C4261, C4263, C4264, C4265, C4266, C4267, C4268, C4269, C4270, C4271, C4272, C4273, C4274, C4278, C4279, C4281, C4311, C4312, C4313, C4314, C4315, C4316, C4318, C4319, C4320, C4321, C4322, C4323, C4324, C4325, C4326, C4328, C4329, C4330, C4331, C4332, C4333, C4334, C4335, C4336, C4337, C4339, C4340, C4341, C4342, C4343, C4344, C4345, C4346, C4347, C4348, C4349, C4350, C4351, C4382, C4383, C4384, C4385, C4386, C4388, C4405, C4406, C4407, C4414, C4415, C4416, C4418, C4421, C4422, C4280), 00763000709648 (C4282, C4300), 00763000709662 (C4306, C4353), 00763000709686 (C4352), 00763000709693 (C4299, C4397), 00763000745462 (C4040), 00763000745486 (C3863, C4069, C4070), 00763000863203 (C4033), 00763000863210 (C3789, C3801, C3804, C3805, C3806, C3807, C3808, C3809, C3878, C3985, C4005, C4006, C4007, C4008, C4009, C4010, C4011), 00763000863241 (C3833, C3890, C3892, C3894, C3980, C3995, C4013, C4029), 00763000863265 (C3826, C4164, C4165), 00763000863272 (C3813, C3814, C3820, C3821, C3827, C3888, C3904, C3906, C3992, C4024, C4026, C4054, C4055, C4056, C4057, C4073, C4112, C4113, C4124, C4147, C4159), 00763000863302 (C3829, C3830, C3831, C3832, C3835, C3837, C3838, C3839, C3840, C3841, C3842, C3843, C3844, C3845, C3848, C3849, C3850, C3851, C3852, C3853, C3854, C3855, C3856, C3857, C3858, C3859, C3860, C3861, C3862, C3864, C3865, C3866, C3867, C3868, C3869, C3870, C3871, C3872, C3873, C3874, C3875, C3876), 00763000871109 (C4127), 00763000871116 (C3989), 00763000871123 (C3986, C3993, C4114), 00763000871130 (C3834, C4058), 00763000967314 (C4389), 00763000967321 (C4296), 00763000987701 (C4364, C4365), 00763000987732 (C4216).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2451-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.