Z-2554-2026Class II
Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser
IPG Medical Corporation / initiated 2026-05-05 / Open, Classified / root cause: design
IPG Medical Corporation began a recall of Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser on . FDA classified it as Class II and gives the reason as "Software anomaly that causes a false display of error code 5018". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2554-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- IPG Medical Corporation
- Product
- Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Software anomaly that causes a false display of error code 5018.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or [email protected].
- Quantity in commerce
- 198 units
- Distribution
- US Nationwide distribution in the state of MA.
- Product codes and lots
- Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2554-2026 | Class II | Ongoing | Voluntary: Firm initiated |