TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300
Howmedica Osteonics Corp. began a recall of TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300 on . FDA classified it as Class II and gives the reason as "Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2461-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Howmedica Osteonics Corp
- Product
- TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Stryker notified consignees on about 05/19/2026 via letter sent using FedEx 2-day service. Consignees were instructed to inform users of the Urgent Medical Device Recall and forward the notice to all individuals who need to be made aware or organizations who have consigned product, immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are on hand, and complete and return the provided Urgent Medical Device Recall Business Reply Form.
- Quantity in commerce
- 7 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, GA, MI, NC, NJ, NY, WV.
- Product codes and lots
- Part Number: 5521-B-300. UDI: (01)07613327026313(17)260921(10) UZD9IB. Lot Number: UZD9IB.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2461-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2462-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY