CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide...
Bard Access Systems, Inc. began a recall of CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement on . FDA classified it as Class II and gives the reason as "Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2658-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Bard Access Systems, Inc.
- Product
- CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement.
- Product code
- LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
- Reason for recall
- Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt
- Quantity in commerce
- 66
- Distribution
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
- Product codes and lots
- CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Lot Number: REKT1554
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2655-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2656-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2657-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2658-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2659-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2660-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2661-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2662-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2663-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2664-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2665-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2666-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2667-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2668-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.