Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2667-2026Class II

CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline...

Bard Access Systems, Inc. / initiated 2026-05-21 / Open, Classified / root cause: manufacturing process

Bard Access Systems, Inc. began a recall of CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code... on . FDA classified it as Class II and gives the reason as "Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2667-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Bard Access Systems, Inc.
Product
CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 S4153108BP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153990 S4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153952 S4153108DCP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741233258 S4153108DGP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154027 S4254108BD Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154140 S4254108BDP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154164 S4254108D Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154119 S4254108DCP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741233234 S4254108DGP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154188 SP4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185052 SP4153108DG Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185069 The Provena" Midline Catheters are intended for short term peripheral access for selected intravenous therapies, blood sampling, and power injection of contrast media.
Product code
PND Midline Catheter
Reason for recall
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt
Quantity in commerce
24,348
Distribution
Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
Product codes and lots
CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 Lot Number: REKN2105 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 Lot Numbers: REKW2429 REKX1484 S4153108BP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153990 Lot Number: REKW2494 S4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741153952 Lot Numbers: REKP0118 REKX2225 S4153108DCP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741233258 Lot Number: REKX3430 S4153108DGP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154027 Lot Numbers: REKQ0118 REKW1477 S4254108BD Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154140 Lot Number: REKY0144 S4254108BDP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154164 Lot Numbers: REKW0447 REKW0953 REKX2214 S4254108D Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154119 Lot Number: REKX3445 S4254108DCP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741233234 Lot Number: REKW2430 S4254108DGP Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154188 Lot Numbers: REKV3192 REKX3437 SP4153108D Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185052 Lot Number: REKW2448 SP4153108DG Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741185069 Lot Number: REKW0959

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2667-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2655-2026Class IIOngoingVoluntary: Firm initiated
Z-2656-2026Class IIOngoingVoluntary: Firm initiated
Z-2657-2026Class IIOngoingVoluntary: Firm initiated
Z-2658-2026Class IIOngoingVoluntary: Firm initiated
Z-2659-2026Class IIOngoingVoluntary: Firm initiated
Z-2660-2026Class IIOngoingVoluntary: Firm initiated
Z-2661-2026Class IIOngoingVoluntary: Firm initiated
Z-2662-2026Class IIOngoingVoluntary: Firm initiated
Z-2663-2026Class IIOngoingVoluntary: Firm initiated
Z-2664-2026Class IIOngoingVoluntary: Firm initiated
Z-2665-2026Class IIOngoingVoluntary: Firm initiated
Z-2666-2026Class IIOngoingVoluntary: Firm initiated
Z-2667-2026Class IIOngoingVoluntary: Firm initiated
Z-2668-2026Class IIOngoingVoluntary: Firm initiated