Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion...
Oscor Inc. began a recall of Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories on . FDA classified it as Class I and gives the reason as "Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2678-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Oscor Inc.
- Product
- Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 10, 2026, Oscor (an Integer subsidiary ) issued a Urgent Medical Device Recall notification to affected consignees. Oscor asked consignees to take the following actions: 1. Product removal is not required, and hospitals may continue to use existing inventory. 2. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 3. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed. 4.Review the information within this letter, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 1) to [email protected]. 5. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 6.If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness.
- Quantity in commerce
- 355 units
- Distribution
- Distribution in the US state of MA
- Product codes and lots
- All Lots within Expiry/GTIN:00885672009793, 00813502010664
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2678-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2679-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2680-2026 | Class I | Ongoing | Voluntary: Firm initiated |