Z-2579-2026Class II

Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions...

Philips North America LLC / initiated 2026-06-05 / Open, Classified / root cause: software

Philips North America Llc began a recall of Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry... on . FDA classified it as Class II and gives the reason as "Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2579-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Root cause tag
software
FDA root cause
Software design
Action
Philips Medical notified consignees on about 06/05/2026 via "URGENT Medical Device Correction" letter. Consignees were notified of the issue and potential hazard and instructed to continue the use of the device but avoid touching the display while the device is booting up and provide this notification to all personnel who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Philips will release new software which will correct this problem. A Philips representative will contact you to arrange for an upgrade to the new software.
Quantity in commerce
909 units
Distribution
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
Product codes and lots
Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2579-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2579-2026Class IIOngoingVoluntary: Firm initiated