Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2674-2026Class II

Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC...

Zimmer Surgical SA / initiated 2026-06-09 / Open, Classified / root cause: packaging

Zimmer Surgical Inc began a recall of Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00... on . FDA classified it as Class II and gives the reason as "Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2674-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer Surgical SA
Product
Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00 Software Version: N/A Product Description: Single Sterile Extension Hose with PLC Connector Component: N/A
Product code
KCY Tourniquet, Pneumatic
Reason for recall
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
On June 8, 2026, the firm began notifying consignees titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to locate and quarantine affected product in their inventory. Customers may either return the affected product, or a Zimmer Biomet sales representative may remove and return the affected product on the customer's behalf. If the product has been further distributed, the distributor should notify downstream customers and accounts of the recall. All customers should ensure that affected personnel are aware of the recall. This recall is related to prior recall RES 98665, conducted to remove Mixing Bowl and Spatula due to similar issues with seal integrity.
Quantity in commerce
14360
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and EMEA.
Product codes and lots
Lot Code: Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271022(10)79701946 ; Box (10-Pack) UDI (01)00889024380455(17)271022(10)79701946 ; Individual UDI (01)00889024376786(17)271022(10)79701946 ; Lot Number 79701946 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271028(10)79701947 ; Box (10-Pack) UDI (01)00889024380455(17)271028(10)79701947 ; Individual UDI (01)00889024376786(17)271028(10)79701947 ; Lot Number 79701947 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271030(10)79701948 ; Box (10-Pack) UDI (01)00889024380455(17)271030(10)79701948 ; Individual UDI (01)00889024376786(17)271030(10)79701948 ; Lot Number 79701948 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280325(10)80555938 ; Box (10-Pack) UDI (01)00889024380455(17)280325(10)80555938 ; Individual UDI (01)00889024376786(17)280325(10)80555938 ; Lot Number 80555938 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280402(10)80555939 ; Box (10-Pack) UDI (01)00889024380455(17)280402(10)80555939 ; Individual UDI (01)00889024376786(17)280402(10)80555939 ; Lot Number 80555939 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280326(10)80555940 ; Box (10-Pack) UDI (01)00889024380455(17)280326(10)80555940 ; Individual UDI (01)00889024376786(17)280326(10)80555940 ; Lot Number 80555940 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280911(10)80581869 ; Box (10-Pack) UDI (01)00889024380455(17)280911(10)80581869 ; Individual UDI (01)00889024376786(17)280911(10)80581869 ; Lot Number 80581869 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280331(10)80581870 ; Box (10-Pack) UDI (01)00889024380455(17)280331(10)80581870 ; Individual UDI (01)00889024376786(17)280331(10)80581870 ; Lot Number 80581870 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280402(10)80765021 ; Box (10-Pack) UDI (01)00889024380455(17)280402(10)80765021 ; Individual UDI (01)00889024376786(17)280402(10)80765021 ; Lot Number 80765021

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2674-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2673-2026Class IIOngoingVoluntary: Firm initiated