Z-2647-2026Class II
RayCare 2024A SP4. Oncology Information Systems
RaySearch Laboratories AB (publ) / initiated 2026-05-20 / Open, Classified
RAYSEARCH LABORATORIES AB began a recall of RayCare 2024A SP4. Oncology Information Systems on . FDA classified it as Class II and gives the reason as "Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2647-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayCare 2024A SP4. Oncology Information Systems.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
- FDA root cause
- Under Investigation by firm
- Action
- A Field Safety Notice, Medical Device Correction #175428 dated 6/2/26 was emailed to consignees. The notice advises consignees to ensure clear verbal or procedural communication when updating the beam set delivery note, avoid making changes from any location other than the control room when treatment is ongoing, avoid relying on beam set delivery note for conveying critical information. Additionally, consignees are to ensure planning staff and users are educated of the provided workaround. Consignees are to verify the software version of their units and reply to the emailed notification. A software update is scheduled for August 2026 to address this issue. Consignees with questions can contact [email protected].
- Quantity in commerce
- 3 units
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- Software Version 8.0.4.60077, UDI 0735000201089120241218, Expiration Date: 2030-01-24.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2646-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2647-2026 | Class II | Ongoing | Voluntary: Firm initiated |