Z-2833-2026Class II
Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150
Stryker Corp. / initiated 2026-07-01 / Open, Classified / root cause: design
Stryker GmbH began a recall of Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150 on . FDA classified it as Class II and gives the reason as "Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2833-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Firm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks should be returned as part of the complaint process.
- Quantity in commerce
- 5,664 units
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, ID, IL, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, SC, TN, TX, VA, WI, WV and the countries of Australia, Austria, Belgium, Canada, Chile, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, UK.
- Product codes and lots
- Catalog Number: 50796150. GTIN: 04546540389954. All lot numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2827-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2828-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2829-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2830-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2831-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2832-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2833-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY