Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number...
CMR SURGICAL LIMITED began a recall of Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1... on . FDA classified it as Class II and gives the reason as "Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2717-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CMR Surgical Inc.
- Product
- Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1 Product Description: The Versius Surgical System is a modular, open console, software controlled, electromechanical system designed for surgeons to perform minimally invasive surgery.
- Product code
- SCV Modular Electromechanical Surgical System
- Reason for recall
- Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- On June 16, 2026, the firm began notifying affected customers. The firm or its local distributor will proactively carry out the secure boot configuration of the affected consoles.
- Quantity in commerce
- 2
- Distribution
- US Nationwide distribution in the state of California.
- Product codes and lots
- UDI-DI 05060548211226 Model No V-SS-1020 Serial Numbers 66495, 66496 Software Revision Version 62.1 There is no expiry date for this device.