Z-2761-2026Class II

Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for...

OLYMPUS Corporation of the Americas / initiated 2026-07-07 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color on . FDA classified it as Class II and gives the reason as "Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2761-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees. Customers were informed of the product issue. While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors. Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed in the customer letter. They should then acknowledge receipt of the letter through the firm's recall web portal. The firm will coordinate repair of the monitors. If customer needs to use the monitor prior to its repair, the firm recommends that a spare unit be available in case issues are encountered with the primary unit during use.
Quantity in commerce
10 units
Distribution
Worldwide - US Nationwide distribution in the states of OH, CA, NJ, FL and the country of Japan.
Product codes and lots
Model Number: OEV321UH. UDI-DI: 04953170418723. Serial Numbers: 7503474, 7503477, 7503479, 7503263, 7503264, 7503262, 7503261, 7503260, 7503259, 7503335.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2761-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2761-2026Class IIOngoingVoluntary: Firm initiated