Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software...
Pride Mobility Products Corp. began a recall of Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product... on . FDA classified it as Class II and gives the reason as "If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2843-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Pride Mobility Products Corp
- Product
- Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
- Product code
- ITI Wheelchair, Powered
- Reason for recall
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
- FDA root cause
- Under Investigation by firm
- Action
- Firm notified dealers and consumers via mailed letters on June 25, 2026. Dealers were instructed to locate affected units and inform consumers of the required software update. Dealers should schedule service appointments with affected consumers. Systems should be updated to Q-Logic software version 2.0.13 or later. The update must be completed following a system power cycle. Consumers were instructed to disable the "Remind Me Later" feature immediately. Consumers should contact their dealer to schedule an appointment for software update.
- Quantity in commerce
- 1102
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
- Product codes and lots
- Lot Code: Q6 Edge 3 Stretto 0060650920018*9 K143383 D246699 41875 ITI
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2841-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2842-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2843-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2844-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2845-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2846-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2847-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2848-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2849-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2850-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2851-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2852-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2853-2026 | Class II | Ongoing | Voluntary: Firm initiated |