AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number...
Windstone Medical Packaging, Inc. began a recall of AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3)... on . FDA classified it as Class II and gives the reason as "The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2802-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Windstone Medical Packaging, Inc.
- Product
- AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Laminectomy Pack, Pack Number AMS13574A; 2) ACDF/Fusion Pack, Pack Number AMS14710B; 3) Spine Pack, Pack Number AMS14991A; 4) Spine Pack, Pack Number AMS14991B; 5) Spinal Pack, Pack Number AMS16278; 6) Spine Pack, Pack Number AMS16499; 7) Spine Pack, Pack Number AMS16709; 8) Lumbar Spine Pack, Pack Number AMS17336; 9) Laminectomy Pack, Pack Number AMS17945; 10) Anterior Cervical Pack, Pack Number AMS17946; 11) Spine Pack, Pack Number AMS3150F; 12) Spine Pack, Pack Number AMS3150G; 13) Spine Pack, Pack Number AMS5451D; 14) ENT Pack, Pack Number AMS9771.
- Product code
- OJH Orthopedic Tray
- Reason for recall
- The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Aligned Medical Solutions (AMS) issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/8/2026 via email. The notice explained the issue, potential risk, and requested the following actions: identify and quarantine affected product, affix the sticker which reads: "RECALL NOTICE Medicom announced the recall of Ritmed Neurological Sponges contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall." " At the time the kit is opened for use any Medicom-Ritmed Neurological Sponges should be identified and set aside. " The recalled Medicom-Ritmed Neurological Sponges should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete the recall reply form and return it via email to: [email protected] even if you have no affected product on hand. Those further distributing the product were directed to notify their distribution partners. For questions contact [email protected].
- Quantity in commerce
- 7306 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, CT, GA, IL, MA, MD, NY, OH, PA, TN, TX, VA, WA.
- Product codes and lots
- 1) Pack Number AMS13574A, UDI B098AMS13574A0, Lot Numbers: 235911, 236897. 2) Pack Number AMS14710B, UDI B098AMS14710B0, Lot Numbers: 224717, 230666, 213914, 236351. 3) Pack Number AMS14991A, UDI B098AMS14991A0, Lot Numbers: 215450, 230108. 4) Pack Number AMS14991B, UDI B098AMS14991B0, Lot Numbers: 233346. 5) Pack Number AMS16278, UDI B098AMS162780, Lot Numbers: 219872, 222999, 226497, 229309, 236465, 237544, 238613. 6) Pack Number AMS16499, UDI B098AMS164990, Lot Numbers: 215711, 218234, 224253, 226168, 226627, 227667, 228446, 230422, 233097, 233399, 233960, 234577, 234907, 236220, 236852, 237358, 238563, 239114, 240373, 241010, 241717. 7) Pack Number AMS16709, UDI B098AMS167090, Lot Numbers: 223361, 225039, 236666, 240344, 241412. 8) Pack Number AMS17336, UDI B098AMS173360, Lot Numbers: 233541, 240967. 9) Pack Number AMS17945, UDI B098AMS179450, Lot Numbers: 235478, 240779, 241565. 10) Pack Number AMS17946, UDI B098AMS179460, Lot Numbers: 235954, 239092, 241514. 11) Pack Number AMS3150F, UDI B098AMS3150F0, Lot Numbers: 211888, 216076, 216357, 218443, 219827, 221799, 223766, 227324, 228151, 229404, 229717, 230623, 231694, 233211, 234586, 234763, 236581, 241332. 12) Pack Number AMS3150G, UDI B098AMS3150G0, Lot Numbers: 242374. 13) Pack Number AMS5451D, UDI B098AMS5451D0, Lot Numbers: 222621, 224568, 238089, 238575, 242113. 14) Pack Number AMS9771, UDI B098AMS97710, Lot Numbers: 176784, 179837, 181956, 184890, 191784, 194017, 196708, 197543, 199137, 201956, 210134, 215434, 217352, 218696, 225332, 226657, 228224, 228678, 230472, 233592, 235614, 236684, 241927.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2799-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2800-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2801-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2802-2026 | Class II | Ongoing | Voluntary: Firm initiated |